Found 15 products
Item#: PN00328 Mfr Part#: 85720 Manufacturer: Leybold
PRICE: $2,700.00
WEB SAVINGS: $701.00
YOU PAY: $1,999.00

Fully Remanufactured Oerlikon Leybold Turbotronik NT 20 Turbo Pump Controller, 85720

 

Leybold Oerlikon NT20 Overview:


The mains voltage of Leybold Turbotronik NT20 is 100/120/220/240 V,%/-15%. Mains frequency of Leybold NT-20 is 50/60 Hz. Interface for data output of Turbotornik NT20 is RS 232/V.24. The weight of Oerlikon NT20 is 7 kg.

The electronic frequency converter TURBOTRONIK NT-20 is used to drive the following turbo molecular pumps:

  • TURBOVAC 150, 360,
  • TURBOVAC 150 CSV, 360 CSV,
  • TURBOVAC 150 V, 360 V,
  • TURBOVAC 151, 151 C, 361, 361 C,
  • TURBOVAC 600, 600 C,
  • TURBOVAC 1000, 1000 C, 1100 C.

These pumps each comprise a three-phase asynchronous motor with the appropriate rating to drive the rotor.

The TURBOTRONIK NT20 converts the single-phase mains voltage into a three-phase AC voltage with regulated frequency and amplitude.

Each installed TURBOVAC has been individually coded. The respective pump code must be entered using the DIP switches provided. For safety reasons, the TURBOVAC will not start up unless the two codes match.

The acceleration sequence, the regulation during operation, and the output speed will vary depending on the installed pump model.

The TURBOTRONIK NT 20 does not support the pump models TURBOVAC 150 H and TURBOVAC 360 H. LEYBOLD customer service can adjust the speed of each TURBOVAC to ± 5%, if required.


THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.

 

Leybold NT 20 Rear Panel:


Leybold NT 20 Rear Panel, NT20, NT-20, 85720

 

Leybold NT 20 Dimensions:


Leybold NT 20 Dimensions, NT20, NT-20, 85720


Click Here for Leybold TURBOTRONIK NT 20 Manual


A&J Vacuum Services offers full repair/service of Leybold Turbotronik NT20 Controller, along with all other brands offered on our website. AJVS also services Leybold Turbo Pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.


Condition: Remanufactured In Stock
Item#: PN18600 Mfr Part#: 800073V0008 Manufacturer: Leybold
PRICE: $852.50
WEB SAVINGS: $77.50
YOU PAY: $775.00

The Oerlikon / Leybold Turbo Drive TD 400 supplies power to the TW series turbo pumps and is used to control their operation.

The Turbo Drive TD 400 is suited for operation of the following pumps:

  • TURBOVAC TW 70 H
  • TURBOVAC TW 220/150 S, TW 220/150/15 S
  • TURBOVAC TW 250 S
  • TURBOVAC TW 300, TW 300 H

Other pumps may only be operated after approval from Leybold or if the operation of such pumps is expressly permitted in the Operating Instructions for the specific pump. The Turbo Drive TD 400 is either integrated in the pump or it is separate and linked to the pump by means of a connecting cable. The Turbo Drive TD 400 requires a supply voltage of 24 V DC. It is equipped with interfaces for programmable controls (REMOTE) and an optional interface for serial communication. The Turbo Drive TD 400 complies with SELV requirements (Safety Extra Low Voltage).

 

THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.

 


Condition: Remanufactured In Stock
Item#: PN00327 Mfr Part#: LNT13 Manufacturer: Leybold
Price upon request

Power consumption

  • briefly (acceleration): up to 64 VA(45 W)
  • continuously (nominal speed): 35 VA(18 W)
  • idle: 6 VA(3 W)

OuTPU ts

  • Voltage 3 x 150 V
  • Nominal frequency 1200 Hz
  • Frequency range of the ouTPU t voltage 220-1250 Hz

Ambient temperatures

  • during operation 0-40°C
  • during storage - 40°C to 85°C

Max. rating of the relay contacts 8 and 10

  • Switching voltage ≤250 V
  • Switching current ≤6 A
  • Switching capacity for DC ≤150 W
  • Switching capacity for AC ≤1500 VA

Max. rating of the relay contacts 4 and 6, mains isolated:  25 Veff / 60 V DC / 1 A
Weight 1.5 kg
Housing dimensions 1/4 19“, 3 HU*



THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.


Condition: Remanufactured Call us for Availability
Item#: PN00329 Mfr Part#: 85572 Manufacturer: Leybold
Price upon request

Overview:

The Leybold Turbotronik NT 150/360 solid-state frequency converter converts the single-phase A.C. voltage of the mains supply into a three-phase A.C. voltage with variable voltage and frequency.

The frequency converter is an externally controlled three-phase transistor-rectifier consisting mainly of six transistor switches. These transistor switches are controlled so that a three-phase connection is formed at the output R, S, T. The diodes connected in antiparallel to the connected consumer into the intermediate D.C. circuit.

After depressing the "START" button, the voltage Uzk in the D.C. intermediate circuit is controlled according to an IU characteristic. Current I and voltage Uzk are proportional to motor current Im and motor voltage Um. The frequency of the three-phase output voltage is varied from a preset minimum value, as a function of the D.C. intermediate circuit voltage Uzk, up to a preset maximum value.

THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.


Condition: Remanufactured Call us for Availability
Item#: PN00330 Mfr Part#: 85719 Manufacturer: Leybold
Price upon request

Leybold NT 151/361  Turbotronik Controller, 230-240 V, 85719

Overview:

The electronic frequency converter TURBOTRONIK NT 151/361 is used to drive the turbomolecular pumps TURBOVAC150, 360, 150 CSV, 360 CSV, 150 V, 360 V,151, 361, 151 C, 361 C.

The TURBOTRONIK NT 151/361 converts the single phase mains voltage into a three-phase AC voltage for driving the asynchronous motor of the connected turbomolecular pump.

The following operational states are indicated by LEDs:

  • Mains voltage applied (POWER, green),
  • Acceleration (ACCEL, yellow),
  • Normal (NORMAL, green),
  • Failure (FAIL, red).

A relay output is also available for the NORMAL signal.

After starting the pump will run up at maximum current. This is indicated by the ACCEL LED.

Having reached approx. 80% of the nominal speed the NORMAL LED comes on and the ACCEL LED goes off.

The motor current is continuously monitored and controlled. If the speed cannot be maintained even at the maximum current due to external influences, for example because of a greater gas supply, the speed will drop but the pump will continue to operate.

When the speed drops below 150 Hz approx. the LED FAIL on the converter lights up.

The outputs are short-circuit proof and protected against no load conditions.

The TURBOTRONIK recognizes which pump is connected.

The connection line between TURBOTRONIK and TURBOVAC may be up to 100 m long.

The TURBOTRONIK is contained within a housing, and can be used as a table top instrument or for installation in a 19� rack. Operation is by means of the START and STOP buttons on the front plate.

THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.

 


Condition: Remanufactured Call us for Availability
Item#: PN07688 Mfr Part#: 85716 Manufacturer: Leybold
Price upon request

Leybold Turbotronik NT 151/361 Controller, 85716

Overview:

The electronic frequency converter TURBOTRONIK NT 151/361 is used to drive the turbomolecular pumps TURBOVAC150, 360, 150 CSV, 360 CSV, 150 V, 360 V,151, 361, 151 C, 361 C.

The TURBOTRONIK NT 151/361 converts the single phase mains voltage into a three-phase AC voltage for driving the asynchronous motor of the connected turbomolecular pump.

The following operational states are indicated by LEDs:

  • Mains voltage applied (POWER, green),
  • Acceleration (ACCEL, yellow),
  • Normal (NORMAL, green),
  • Failure (FAIL, red).

A relay output is also available for the NORMAL signal.

After starting the pump will run up at maximum current. This is indicated by the ACCEL LED.

Having reached approx. 80% of the nominal speed the NORMAL LED comes on and the ACCEL LED goes off.

The motor current is continuously monitored and controlled. If the speed cannot be maintained even at the maximum current due to external influences, for example because of a greater gas supply, the speed will drop but the pump will continue to operate.

When the speed drops below 150 Hz approx. the LED FAIL on the converter lights up.

The outputs are short-circuit proof and protected against no load conditions.

The TURBOTRONIK recognizes which pump is connected.

The connection line between TURBOTRONIK and TURBOVAC may be up to 100 m long.

The TURBOTRONIK is contained within a housing, and can be used as a table top instrument or for installation in a 19 inch rack. Operation is by means of the START and STOP buttons on the front plate.

THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.

 


Condition: Remanufactured Call us for Availability
Item#: PN07122 Mfr Part#: 85562 Manufacturer: Leybold
Price upon request

Fully Remanufactured Oerlikon Leybold Turbotronik NT 20 Turbo Pump Controller, 85562

 

Leybold Oerlikon NT20 Overview:


The mains voltage of Leybold Turbotronik NT20 is 100/120/220/240 V,%/-15%. Mains frequency of Leybold NT-20 is 50/60 Hz. Interface for data output of Turbotornik NT20 is RS 232/V.24. The weight of Oerlikon NT20 is 7 kg.

The electronic frequency converter TURBOTRONIK NT-20 is used to drive the following turbo molecular pumps:

  • TURBOVAC 150, 360,
  • TURBOVAC 150 CSV, 360 CSV,
  • TURBOVAC 150 V, 360 V,
  • TURBOVAC 151, 151 C, 361, 361 C,
  • TURBOVAC 600, 600 C,
  • TURBOVAC 1000, 1000 C, 1100 C.

These pumps each comprise a three-phase asynchronous motor with the appropriate rating to drive the rotor.

The TURBOTRONIK NT20 converts the single-phase mains voltage into a three-phase AC voltage with regulated frequency and amplitude.

Each installed TURBOVAC has been individually coded. The respective pump code must be entered using the DIP switches provided. For safety reasons, the TURBOVAC will not start up unless the two codes match.

The acceleration sequence, the regulation during operation, and the output speed will vary depending on the installed pump model.

The TURBOTRONIK NT 20 does not support the pump models TURBOVAC 150 H and TURBOVAC 360 H. LEYBOLD customer service can adjust the speed of each TURBOVAC to ± 5%, if required.


THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.

 

Leybold NT 20 Rear Panel:


Leybold NT 20 Rear Panel, NT20, NT-20, 85562

 

Leybold NT 20 Dimensions:


Leybold NT 20 Dimensions, NT20, NT-20, 85562


Click Here for Leybold TURBOTRONIK NT 20 Manual


A&J Vacuum Services offers full repair/service of Leybold Turbotronik NT20 Controller, along with all other brands offered on our website. AJVS also services Leybold Turbo Pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.


Condition: Remanufactured Call us for Availability
Item#: PN07820 Mfr Part#: 11366 Manufacturer: Leybold
Price upon request

 

Overview:

The Oerlikon / Leybold Turbo Drive S Controller supplies power to the TW series turbomolecular pumps and is used to control their operation.

The Turbo Drive S is suited for operation of the following pumps:

  • TURBOVAC TW 70 H
  • TURBOVAC TW 220/150 S, TW 220/150/15 S
  • TURBOVAC TW 250 S
  • TURBOVAC TW 300, TW 300 H

The Turbo Drive S is either integrated in the pump or it is separate and linked to the pump by means of a connecting cable.

The Turbo Drive S requires a supply voltage of 24 V DC. It is equipped with interfaces for programmable controls (REMOTE) and an optional interface for serial communication.

 

THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.

 


Condition: Remanufactured Call us for Availability
Item#: PN07123 Mfr Part#: 85721 Manufacturer: Leybold
Price upon request

Fully Remanufactured Oerlikon Leybold Turbotronik NT 20 Turbo Pump Controller, 85721

 

Leybold Oerlikon NT20 Overview:


The mains voltage of Leybold Turbotronik NT20 is 100/120/220/240 V,%/-15%. Mains frequency of Leybold NT-20 is 50/60 Hz. Interface for data output of Turbotornik NT20 is RS 232/V.24. The weight of Oerlikon NT20 is 7 kg.

The electronic frequency converter TURBOTRONIK NT-20 is used to drive the following turbo molecular pumps:

  • TURBOVAC 150, 360,
  • TURBOVAC 150 CSV, 360 CSV,
  • TURBOVAC 150 V, 360 V,
  • TURBOVAC 151, 151 C, 361, 361 C,
  • TURBOVAC 600, 600 C,
  • TURBOVAC 1000, 1000 C, 1100 C.

These pumps each comprise a three-phase asynchronous motor with the appropriate rating to drive the rotor.

The TURBOTRONIK NT20 converts the single-phase mains voltage into a three-phase AC voltage with regulated frequency and amplitude.

Each installed TURBOVAC has been individually coded. The respective pump code must be entered using the DIP switches provided. For safety reasons, the TURBOVAC will not start up unless the two codes match.

The acceleration sequence, the regulation during operation, and the output speed will vary depending on the installed pump model.

The TURBOTRONIK NT 20 does not support the pump models TURBOVAC 150 H and TURBOVAC 360 H. LEYBOLD customer service can adjust the speed of each TURBOVAC to ± 5%, if required.


THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.

 

Leybold NT 20 Rear Panel:


Leybold NT 20 Rear Panel, NT20, NT-20, 85721

 

Leybold NT 20 Dimensions:


Leybold NT 20 Dimensions, NT20, NT-20, 85721


Click Here for Leybold TURBOTRONIK NT 20 Manual


A&J Vacuum Services offers full repair/service of Leybold Turbotronik NT20 Controller, along with all other brands offered on our website. AJVS also services Leybold Turbo Pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.


Condition: Remanufactured Call us for Availability
Item#: PN07837 Mfr Part#: 85904 Manufacturer: Leybold
Price upon request

 

Overview:

The TurboTronik NT 10, NT 12 and NT 13 frequency converters are used to drive TURBOVAC 50 and 50 D.

The electronic frequency converters convert the single-phase supply voltage into a three-phase AC voltage to control the TURBOVAC induction motor.

The frequency converters operate according to a specific cycle: The motor is driven for approximately 1 second and the monitoring measurements are executed in the following 0.1 seconds.

The following operating statuses are displayed using LEDs:

  • supply voltage present
  • run-up
  • standard and
  • fault Optionally

Various signals are available via 2 relay outputs.

After the start, the pump continuously accelerates with the maximum current. This is identified by the yellow LED "ACCEL" (acceleration) which is lit. The green "NORMAL" LED is lit and the "ACCEL" LED goes dark when approximately 90% of the rated speed is reached.

The speed is continuously monitored and controlled. When the speed cannot be maintained, even at the maximum current, as a result of external influences, e.g. excessive gas feed, the speed decreases and the pump continues to operate.

When a rotational frequency of approximately 500 Hz is fallen below, the "ACCEL" LED lights-up on the frequency converter. The TURBOTRONIK now attempts to accelerate the pump up to its reference frequency.

The frequency converter outputs are no-load- and short-circuit proof.

The electronic TURBOTRONIK frequency converters can be connected to external open-loop control- and monitoring devices with electrical isolation.

The cable connections between TURBOTRONIK - TURBOVAC can be a maximum of 100 m.

 

THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.

 


Condition: Remanufactured Call us for Availability
Contact
address
A&J Vacuum Services
16 Somerset Place, Clifton, NJ 07012
customer service
Customer Service
Phone: 973-249-0854, Fax: 973-249-0855
working hours
Service Hours
8:00am to 5:00pm, Monday thru Friday EST
All trademarks, trade names or logons mentioned or used are the property of their respective owners.