Oerlikon / Leybold PhoeniXL 300 Dry Leak Detector With ScrollVac 30 D Dry Scroll Pump, 250001
Item#: PN15878 Mfr Part#: 251001 Manufacturer: Leybold
Condition: Remanufactured Call us for Availability
Remanufactured Oerlikon / Leybold PhoeniXL 300 Dry Leak Detector With ScrollVac 30D Dry Scroll Vacuum Pump, 250001

Features:
  • Max. Inlet Pressure: 15 mbar
  • Measurement Range: 11 decades
  • Inlet Port: DN 25 KF
  • Power Consumption: 350 VA

Price upon request
Item#: PN15878 Mfr Part#: 251001
Oerlikon / Leybold PhoeniXL 300 Dry Leak Detector With ScrollVac 30 D Dry Scroll Pump, 250001
Item#: PN15878 Mfr Part#: 251001 Manufacturer: Leybold

Oerlikon / Leybold PhoeniXL 300 Dry Leak Detector With ScrollVac 30D Dry Scroll Vacuum Pump, 250001

 

Oerlikon / Leybold 300 Overview:


The PhoeniXL is a leak detector for Helium or Hydrogen. This instrument may be used to detect the location and the size of leaks on objects under test in two different ways:

  • When the test sample has been evacuated first and is sprayed with helium on the outside. It is required that a vacuum connection is provided between the PhoeniXL and the test sample (vacuum mode).
  • When a helium overpressure is provided in the test sample and the test sample is searched from the outside with a sniffer probe which is attached to the inlet port (sniffer mode).

The PhoeniXL is to be used for leak detection only. It must not be used as a pumping system (esp. pumping aggressive or humid gases.)

Pumping condensable gases and steams: When pumping test sample water vapour that is inside the test object can attain to the fore pump. With the water vapor that is in the air - especially in humid areas or when using humid or wet test samples - the acceptable compatibility of water vapor or capacity of water vapor respectively can be exceeded. The steam in the oil of the pump condeses when the water vapor rises over the acceptable value. So the attribute of the oil changes and danger of corrosion occures for the pump.

While using the leak detector with condensable gases and steams the oil of the fore pump has to be controlled regularly. So you can recognize a condensation of water vapor in the pump. Usually the oil is light and lucent. When water vapor is inside it gets blear and milky at operating state temperature. When turning the pump off water vapor condensates and raises the part of water in the oil.

The leak detector must not directly be switched off after the process, in which condensable gases or steams are pumped, is finished. It must be running (at least 20 minutes) with opend gas ballast valve until the oil of the pump is freed from detached steam.

 

Oerlikon / Leybold 300 Front Panel:


Oerlikon / Leybold 300 PhoeniXL, 251001

 

Oerlikon / Leybold 300 Dimensions:


Oerlikon / Leybold 300 PhoeniXL, 251001

 


Click Here for Leybold PhoeniXL 300 Operating Instruction

Click Here for Leybold Oerlikon Helium Leak Detectors

Click Here for Leybold Leybold Product Brochure


A&J Vacuum Services offers full repair/service of Oerlikon / Leybold PhoeniXL 300, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.

This item is non-returnable/non-refundable

Oerlikon / Leybold PhoeniXL 300 Dry Leak Detector With ScrollVac 30D Dry Scroll Vacuum Pump, 250001

 

Oerlikon / Leybold 300 Technical Data:


Max. Inlet Pressure: 15 mbar
Minimum Detectable Helium Leak Rates:

Vacuum Mode: < 3 x 10-11 mbar l/s

 

Sniffer Mode: < 1 x 10-7 mbar l/s

Minimum Detectable Hydrogen Leak Rates:

Vacuum Mode: < 1 x 10-8 mbar l/s

 

Sniffer Mode: < 1 x 10-7 mbar l/s

Measurement Range: 11 decades
Pumping Speed (Helium) At Inlet:

Gross: .02 l/s

 

Precision: .4 l/s

 

Fine: > 2.5 l/s

Inlet Port: DN 25 KF
Power Consumption: 350 VA

Click Here for Leybold PhoeniXL 300 Operating Instruction

Click Here for Leybold Oerlikon Helium Leak Detectors

Click Here for Leybold Leybold Product Brochure


A&J Vacuum Services offers full repair/service of Oerlikon / Leybold PhoeniXL 300, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.

| Service and Repair
We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.

Overview

Oerlikon / Leybold PhoeniXL 300 Dry Leak Detector With ScrollVac 30 D Dry Scroll Pump, 250001
Item#: PN15878 Mfr Part#: 251001 Manufacturer: Leybold

Oerlikon / Leybold PhoeniXL 300 Dry Leak Detector With ScrollVac 30D Dry Scroll Vacuum Pump, 250001

 

Oerlikon / Leybold 300 Overview:


The PhoeniXL is a leak detector for Helium or Hydrogen. This instrument may be used to detect the location and the size of leaks on objects under test in two different ways:

  • When the test sample has been evacuated first and is sprayed with helium on the outside. It is required that a vacuum connection is provided between the PhoeniXL and the test sample (vacuum mode).
  • When a helium overpressure is provided in the test sample and the test sample is searched from the outside with a sniffer probe which is attached to the inlet port (sniffer mode).

The PhoeniXL is to be used for leak detection only. It must not be used as a pumping system (esp. pumping aggressive or humid gases.)

Pumping condensable gases and steams: When pumping test sample water vapour that is inside the test object can attain to the fore pump. With the water vapor that is in the air - especially in humid areas or when using humid or wet test samples - the acceptable compatibility of water vapor or capacity of water vapor respectively can be exceeded. The steam in the oil of the pump condeses when the water vapor rises over the acceptable value. So the attribute of the oil changes and danger of corrosion occures for the pump.

While using the leak detector with condensable gases and steams the oil of the fore pump has to be controlled regularly. So you can recognize a condensation of water vapor in the pump. Usually the oil is light and lucent. When water vapor is inside it gets blear and milky at operating state temperature. When turning the pump off water vapor condensates and raises the part of water in the oil.

The leak detector must not directly be switched off after the process, in which condensable gases or steams are pumped, is finished. It must be running (at least 20 minutes) with opend gas ballast valve until the oil of the pump is freed from detached steam.

 

Oerlikon / Leybold 300 Front Panel:


Oerlikon / Leybold 300 PhoeniXL, 251001

 

Oerlikon / Leybold 300 Dimensions:


Oerlikon / Leybold 300 PhoeniXL, 251001

 


Click Here for Leybold PhoeniXL 300 Operating Instruction

Click Here for Leybold Oerlikon Helium Leak Detectors

Click Here for Leybold Leybold Product Brochure


A&J Vacuum Services offers full repair/service of Oerlikon / Leybold PhoeniXL 300, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.

Oerlikon / Leybold PhoeniXL 300 Dry Leak Detector With ScrollVac 30D Dry Scroll Vacuum Pump, 250001

 

Oerlikon / Leybold 300 Technical Data:


Max. Inlet Pressure: 15 mbar
Minimum Detectable Helium Leak Rates:

Vacuum Mode: < 3 x 10-11 mbar l/s

 

Sniffer Mode: < 1 x 10-7 mbar l/s

Minimum Detectable Hydrogen Leak Rates:

Vacuum Mode: < 1 x 10-8 mbar l/s

 

Sniffer Mode: < 1 x 10-7 mbar l/s

Measurement Range: 11 decades
Pumping Speed (Helium) At Inlet:

Gross: .02 l/s

 

Precision: .4 l/s

 

Fine: > 2.5 l/s

Inlet Port: DN 25 KF
Power Consumption: 350 VA

Click Here for Leybold PhoeniXL 300 Operating Instruction

Click Here for Leybold Oerlikon Helium Leak Detectors

Click Here for Leybold Leybold Product Brochure


A&J Vacuum Services offers full repair/service of Oerlikon / Leybold PhoeniXL 300, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.

We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.


Contact
address
A&J Vacuum Services
16 Somerset Place, Clifton, NJ 07012
customer service
Customer Service
Phone: 973-249-0854, Fax: 973-249-0855
working hours
Service Hours
8:00am to 5:00pm, Monday thru Friday EST
All trademarks, trade names or logons mentioned or used are the property of their respective owners.