Edwards APG M NW25 Active Pirani Vacuum Gauge, Stainless Steel, D02172000
Item#: PN16000 Mfr Part#: D02172000 Manufacturer: Edwards Vacuum
Condition: New
New Edwards APG M NW25 Active Pirani Vacuum Gauge, Stainless Steel, D02172000

Features:
  • Pressure Range: Atmosphere to 10-3 mbar
  • Maximum Over Pressure: 6 bar absolute
  • Output Signal: 2 to 10 V dc
  • Maximum over pressure 6 bar absolute

PRICE: $1,120.00
WEB SAVINGS: $130.00
YOU PAY: $990.00
Item#: PN16000 Mfr Part#: D02172000
Edwards APG M NW25 Active Pirani Vacuum Gauge, Stainless Steel, D02172000
Item#: PN16000 Mfr Part#: D02172000 Manufacturer: Edwards Vacuum

Edwards APG M NW25 Active Pirani Vacuum Gauge, Stainless Steel, D02172000

 

APG M NW25 Overview:


The Edwards APG, is a Pirani vacuum gauge and gauge controller in a single compact unit. The measurement range of the APG-M and of the corrosion-resistant APG-MP is 100 to 1 x 10-3 mbar: the range of the APG-L is 10 to 1 x 10-4 mbar. All of the gauges will indicate pressures up to 1000 mbar at reduced accuracy. The APG requires a 20 to 36 V d.c. power supply: it has a 2 to 10 V d.c. analogue output which is related to pressure. The APG is compatible with all of the Edwards AGC’s (Active Gauge Controllers) and with the appropriate versions of the Edwards AGD’s (Active Gauge Displays). Alternatively, you can use an independent power supply for the APG and you can read the Edwards APG output signal with a voltmeter or an analogue- to-digital converter.

An 8-way electrical connector socket on the APG is used to connect the APG to your AGC, AGD or electrical supply and voltmeter. Electrical cables fitted with suitable connector plugs are available as accessories. A gauge identification signal is available on the electrical connector: this signal is used by Edwards AGC’s to identify which type of Active gauge is connected, that is one of the following types:

  • APG-M/MP
  • APG-L

 

*THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED*

 


A&J Vacuum services and repairs Edwards Vacuum Measurement Transmitters, along with all other Edwards Vacuum and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

This item is non-returnable/non-refundable

Edwards APG M NW25 Active Pirani Vacuum Gauge, Stainless Steel, D02172000

 

APG M NW25 Technical Data:


Pressure range  
    APG-M Atmosphere to 10-3 mbar
    APG-L Atmosphere to 10-4 mbar
Maximum over pressure 6 bar absolute
Power supply 20 to 36 V dc (max 1 V ripple)
Power consumption 1.5 W max
Output signal 2 to 10 V dc
Adjustments Set vacuum and set atmosphere
Set-point  
    Range of set-point 5% to 95% full-scale voltage
    Fixed hysterisis 7% full-scale voltage
    Level setting resolution ±2% full-scale voltage
Rating 40 V dc, 100mA max
Temperature range  
    Operating 5 to 60 °C
    Storage 0 to 70 °C
Materials exposed to vacuum  
Filament  
    APG-M, APG-L Gold-coated tungsten
    APG-MP Platinum/rhodium (90/10)
    Other materials Aluminium HETF, nickel, fluoroelas- tomer, PTFE, stainless steel 316
Internal volume NW: 6cm3 15mm O.D.: 10cm3
Weight          
    Aluminium 120g
    Stainless steel 200g
Socket connector 8 way FCC68 compatible
Vacuum fitting tube NW 16 or NW 25 or 15 mm O.D.
Standards  
    Electronic design EN 61010-1
    Electrical noise immunity EN 50081-1, EN 50082-1
Flame-retardant casing UL94

 


A&J Vacuum services and repairs Edwards Vacuum Measurement Transmitters, along with all other Edwards Vacuum and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Name
Price
Item#:
PN19714
Mfr Part#:
D40001010
$49.99
Item#:
PN19715
Mfr Part#:
D40001030
$89.95
Item#:
PN19716
Mfr Part#:
D40001050
$115.95
Item#:
PN19717
Mfr Part#:
D40001100
$179.00
Item#:
PN19718
Mfr Part#:
D40001150
$228.00
Item#:
PN19719
Mfr Part#:
D40001250
$305.00
Item#:
PN19720
Mfr Part#:
D40001500
$725.00
Item#:
PN19721
Mfr Part#:
D40001999
$720.00
| Service and Repair
We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.

Overview

Edwards APG M NW25 Active Pirani Vacuum Gauge, Stainless Steel, D02172000
Item#: PN16000 Mfr Part#: D02172000 Manufacturer: Edwards Vacuum

Edwards APG M NW25 Active Pirani Vacuum Gauge, Stainless Steel, D02172000

 

APG M NW25 Overview:


The Edwards APG, is a Pirani vacuum gauge and gauge controller in a single compact unit. The measurement range of the APG-M and of the corrosion-resistant APG-MP is 100 to 1 x 10-3 mbar: the range of the APG-L is 10 to 1 x 10-4 mbar. All of the gauges will indicate pressures up to 1000 mbar at reduced accuracy. The APG requires a 20 to 36 V d.c. power supply: it has a 2 to 10 V d.c. analogue output which is related to pressure. The APG is compatible with all of the Edwards AGC’s (Active Gauge Controllers) and with the appropriate versions of the Edwards AGD’s (Active Gauge Displays). Alternatively, you can use an independent power supply for the APG and you can read the Edwards APG output signal with a voltmeter or an analogue- to-digital converter.

An 8-way electrical connector socket on the APG is used to connect the APG to your AGC, AGD or electrical supply and voltmeter. Electrical cables fitted with suitable connector plugs are available as accessories. A gauge identification signal is available on the electrical connector: this signal is used by Edwards AGC’s to identify which type of Active gauge is connected, that is one of the following types:

  • APG-M/MP
  • APG-L

 

*THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED*

 


A&J Vacuum services and repairs Edwards Vacuum Measurement Transmitters, along with all other Edwards Vacuum and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Edwards APG M NW25 Active Pirani Vacuum Gauge, Stainless Steel, D02172000

 

APG M NW25 Technical Data:


Pressure range  
    APG-M Atmosphere to 10-3 mbar
    APG-L Atmosphere to 10-4 mbar
Maximum over pressure 6 bar absolute
Power supply 20 to 36 V dc (max 1 V ripple)
Power consumption 1.5 W max
Output signal 2 to 10 V dc
Adjustments Set vacuum and set atmosphere
Set-point  
    Range of set-point 5% to 95% full-scale voltage
    Fixed hysterisis 7% full-scale voltage
    Level setting resolution ±2% full-scale voltage
Rating 40 V dc, 100mA max
Temperature range  
    Operating 5 to 60 °C
    Storage 0 to 70 °C
Materials exposed to vacuum  
Filament  
    APG-M, APG-L Gold-coated tungsten
    APG-MP Platinum/rhodium (90/10)
    Other materials Aluminium HETF, nickel, fluoroelas- tomer, PTFE, stainless steel 316
Internal volume NW: 6cm3 15mm O.D.: 10cm3
Weight          
    Aluminium 120g
    Stainless steel 200g
Socket connector 8 way FCC68 compatible
Vacuum fitting tube NW 16 or NW 25 or 15 mm O.D.
Standards  
    Electronic design EN 61010-1
    Electrical noise immunity EN 50081-1, EN 50082-1
Flame-retardant casing UL94

 


A&J Vacuum services and repairs Edwards Vacuum Measurement Transmitters, along with all other Edwards Vacuum and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Name
Price
Item#:
PN19714
Mfr Part#:
D40001010
$49.99
Item#:
PN19715
Mfr Part#:
D40001030
$89.95
Item#:
PN19716
Mfr Part#:
D40001050
$115.95
Item#:
PN19717
Mfr Part#:
D40001100
$179.00
Item#:
PN19718
Mfr Part#:
D40001150
$228.00
Item#:
PN19719
Mfr Part#:
D40001250
$305.00
Item#:
PN19720
Mfr Part#:
D40001500
$725.00
Item#:
PN19721
Mfr Part#:
D40001999
$720.00
We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.


Contact
address
A&J Vacuum Services
16 Somerset Place, Clifton, NJ 07012
customer service
Customer Service
Phone: 973-249-0854, Fax: 973-249-0855
working hours
Service Hours
8:00am to 5:00pm, Monday thru Friday EST
All trademarks, trade names or logons mentioned or used are the property of their respective owners.