The Electrode assembly kit contains one of each of the following components:
Technology
The AIM-X-NW25 Gauge is an inverted magnetron gauge head and gauge controller in a single compact unit. The gauge operates as a cold cathode ionisation gauge, in which pressure is measured indirectly as a function of the current that flows in a Townsend discharge maintained in the body tube.
The measurement range of the AIM-X-NW25 Gauge is 1 x 10-9 to 1 x 10-2 mbar.
The AIM Gauge requires a 13.5 to 36 V d.c. power supply: it has a 2 to 10 V d.c. analogue output which is related to pressure.
The AIM-X-NW25 Gauge is compatible with all of Edwards Active Gauge Controllers and with the appropriate versions of the Active Gauge Displays. Alternatively you can use an independent power supply for the AIM Gauge and you can read the AIM Gauge output signal with a voltmeter or an analogue-to-digital converter.
Applications
All vacuum applications requiring rugged and reliable pressure indication in the range 1 x 10-2 to 1 x 10-9 mbar. Typical processes range from general vacuum through industrial coaters and furnaces to scientific instruments and semicon.
*THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED*
Operating Temperature; +5 to +60 *C
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.
The Electrode assembly kit contains one of each of the following components:
Technology
The AIM-X-NW25 Gauge is an inverted magnetron gauge head and gauge controller in a single compact unit. The gauge operates as a cold cathode ionisation gauge, in which pressure is measured indirectly as a function of the current that flows in a Townsend discharge maintained in the body tube.
The measurement range of the AIM-X-NW25 Gauge is 1 x 10-9 to 1 x 10-2 mbar.
The AIM Gauge requires a 13.5 to 36 V d.c. power supply: it has a 2 to 10 V d.c. analogue output which is related to pressure.
The AIM-X-NW25 Gauge is compatible with all of Edwards Active Gauge Controllers and with the appropriate versions of the Active Gauge Displays. Alternatively you can use an independent power supply for the AIM Gauge and you can read the AIM Gauge output signal with a voltmeter or an analogue-to-digital converter.
Applications
All vacuum applications requiring rugged and reliable pressure indication in the range 1 x 10-2 to 1 x 10-9 mbar. Typical processes range from general vacuum through industrial coaters and furnaces to scientific instruments and semicon.
*THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED*
Operating Temperature; +5 to +60 *C
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.