Edwards TIC Instrument Controller 200W, RS232/485, D39722000, With Support of Turbo Pump and 3 Active Vacuum Gauges
Item#: PN24546 Mfr Part#: D39722000 Manufacturer: Edwards Vacuum
Condition: New In Stock
New Edwards TIC Instrument Controller 200W, RS232/485, D39722000, With Support of Turbo Pump and 3 Active Vacuum Gauges

Features:
  • TIC Turbo Pump Controller
  • P/N D39722000
  • With 3 Gauge Controller
  • Support 1 Turbo Pump and Backing Pump

PRICE: $3,925.00
WEB SAVINGS: $335.00
YOU PAY: $3,590.00
Item#: PN24546 Mfr Part#: D39722000
Edwards TIC Instrument Controller 200W, RS232/485, D39722000, With Support of Turbo Pump and 3 Active Vacuum Gauges
Item#: PN24546 Mfr Part#: D39722000 Manufacturer: Edwards Vacuum

Edwards TIC Instrument Controller 200W, RS232/485, D39722000, With Support of Turbo Pump, Backing Pump and 3 Active Vacuum Gauges

Edwards TIC D39722000 Turbo Pump and Instrument Controller Overview:


The Edwards compact turbo controller with a large clear graphical display, an intuitive user interface and serial communications providing full remote control and data logging functions via a new Windows™ based PC program. The controller automatically recognizes and supports one 24 V turbomolecular pump from the EXT/nEXT range

Edwards Turbo Pump cooling and vent valve support is provided directly from the controller. Mains backing pumps (up to nXDS20i/RV12) may be controlled via an optional relay box. 2 different power variants are available, 100W or 200W which determines the ramp speed of the turbo pump.

In addition 200W models have the ability to power a 24Vdc backing pump such as our XDD1 or nXDS through a 15-way D. Turbo and instrument models add the ability to control up to 3 of our range of Active gauges (excluding IGC & Barocels).

Edwards TTI D39722000 Instrument Features:


  • TIC software is structured through a series of easily accessible screens, similar to a mobile telephone.
  • Supports 1 Turbo Pump (EXT75DX / nEXT85 / nEXT240 / nEXT300 / nEXT400)
  • TIC automatically recognizes which turbo is connected and displays the appropriate information on the default summary screen. Scrolling and selecting accesses the control and set-up menus for that item.
  • Edwards Turbo speed is shown as a proportion of full speed by a bar graph. ‘Norm’ indicates that the turbo has reached its Normal Speed, whilst forward or reverse facing chevrons indicate acceleration and deceleration.
  • Instrument supporting TIC variants recognize and control the following gauges: APG, APGX, ATC, ASG, AIM, WRG & AIGX.  No more than one AIGX gauge may be connected to TIC at a time.
  • Backing pump 24 V (For mains backing pump support, see below.)
    200 W turbo supporting TIC variants recognize and control the following 24 V backing pumps: XDD1
  • Relay box (optional)
    A range of relay boxes has been developed to allow TIC to operate three 250 V a.c. 3 A (non-inductive) changeover relays, mains backing pumps and accessories.
  • The changeover relays are activated by the TIC set point outputs. The mains backing pump relay controls a backing line isolation valve, such that when the backing pump is switched off the isolation valve closes.
    The relay box is connected to the TIC via the logic interface connector, which is also provided with a bypass connector for interfacing with OEM equipment.
  • Pumps status is displayed, giving a clear indication of what is currently happening in the vacuum system.
  • In the event of an error occurring, Edwards TIC will display either a WARNING or flash an ALARM. A warning advises of a condition outside normal parameters, requiring no action, but an alarm must be cleared before normal operation may resume.

 

Relay box options

  • TIC Relay Box Instruments
  • Changeover relays (3 off 250 V a.c. 3 A)
  • TIC Relay Box Small Backing pump
  • Heater band & backing line isolation valve
  • Small backing pump
  • TIC Relay Box Inst & Small Backing pump
  • Changeover relays (3 off 250 V a.c. 3 A)
  • Heater band & backing line isolation valve
  • Small backing pump
  • Compatible mains backing pumps and accessories
    • Small backing pumps: E2M0.7 & 1.5, RV3, 5, 8 & 12, XDS5C & 10C, XDS5 & 10 & ESDP12
    • Heater band: BX70 & BX250
    • Backing line isolation valve: LCPV16EKA & LCPV25EKA

 


A&J Vacuum services and repairs Edwards Tic Instrument, along with all other Edwards Controller and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Edwards TIC Instrument Controller 200W, RS232/485, D39722000, With Support of Turbo Pump, Backing Pump and 3 Active Vacuum Gauges

Edwards TTIC D39722000 Instrument Technical Data:


Pump / TIC power 100W / 200W
  EXT75DX Fast / Fast
  nEXT 80W Slow / Slow
  nEXT 160W Slow / Fast
Mains input  
    Electrical supply 90 to 264 V a.c. 47 to 63 Hz
  Power consumption (max) 215 VA
  Peak inrush current 10.3 A @ 110 V a.c. / 23.0 A @ 230 V a.c.
  Earth stud M4
Auxiliary terminals  
  Air cooling fan 24 V d.c. 3 W max, ACX70, ACX75 & ACX250H
  Vent valve 24 V d.c. 2 W max, TAV5 & TAV6
Dimensions  
  Electronics housing 110 mm high x 105 mm wide x 245 mm deep
  Front panel 106 mm wide x 128 mm high
Weight 3.5 kg
Operating temp +0 ° to +40 °C
Storage temp -30 ° to +70 °C
Max ambient operating humidity 90% RH non-condensing at 40 °C
Max operating altitude 3000 m
Electronic Design EN 61010-1
Electromagnetic Compatibility EN 61326 Industrial Location, Class B Emissions
Enclosure rating IP20

 


A&J Vacuum services and repairs Edwards Tic Instrument, along with all other Edwards Controller and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

| Service and Repair
We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.

Overview

Edwards TIC Instrument Controller 200W, RS232/485, D39722000, With Support of Turbo Pump and 3 Active Vacuum Gauges
Item#: PN24546 Mfr Part#: D39722000 Manufacturer: Edwards Vacuum

Edwards TIC Instrument Controller 200W, RS232/485, D39722000, With Support of Turbo Pump, Backing Pump and 3 Active Vacuum Gauges

Edwards TIC D39722000 Turbo Pump and Instrument Controller Overview:


The Edwards compact turbo controller with a large clear graphical display, an intuitive user interface and serial communications providing full remote control and data logging functions via a new Windows™ based PC program. The controller automatically recognizes and supports one 24 V turbomolecular pump from the EXT/nEXT range

Edwards Turbo Pump cooling and vent valve support is provided directly from the controller. Mains backing pumps (up to nXDS20i/RV12) may be controlled via an optional relay box. 2 different power variants are available, 100W or 200W which determines the ramp speed of the turbo pump.

In addition 200W models have the ability to power a 24Vdc backing pump such as our XDD1 or nXDS through a 15-way D. Turbo and instrument models add the ability to control up to 3 of our range of Active gauges (excluding IGC & Barocels).

Edwards TTI D39722000 Instrument Features:


  • TIC software is structured through a series of easily accessible screens, similar to a mobile telephone.
  • Supports 1 Turbo Pump (EXT75DX / nEXT85 / nEXT240 / nEXT300 / nEXT400)
  • TIC automatically recognizes which turbo is connected and displays the appropriate information on the default summary screen. Scrolling and selecting accesses the control and set-up menus for that item.
  • Edwards Turbo speed is shown as a proportion of full speed by a bar graph. ‘Norm’ indicates that the turbo has reached its Normal Speed, whilst forward or reverse facing chevrons indicate acceleration and deceleration.
  • Instrument supporting TIC variants recognize and control the following gauges: APG, APGX, ATC, ASG, AIM, WRG & AIGX.  No more than one AIGX gauge may be connected to TIC at a time.
  • Backing pump 24 V (For mains backing pump support, see below.)
    200 W turbo supporting TIC variants recognize and control the following 24 V backing pumps: XDD1
  • Relay box (optional)
    A range of relay boxes has been developed to allow TIC to operate three 250 V a.c. 3 A (non-inductive) changeover relays, mains backing pumps and accessories.
  • The changeover relays are activated by the TIC set point outputs. The mains backing pump relay controls a backing line isolation valve, such that when the backing pump is switched off the isolation valve closes.
    The relay box is connected to the TIC via the logic interface connector, which is also provided with a bypass connector for interfacing with OEM equipment.
  • Pumps status is displayed, giving a clear indication of what is currently happening in the vacuum system.
  • In the event of an error occurring, Edwards TIC will display either a WARNING or flash an ALARM. A warning advises of a condition outside normal parameters, requiring no action, but an alarm must be cleared before normal operation may resume.

 

Relay box options

  • TIC Relay Box Instruments
  • Changeover relays (3 off 250 V a.c. 3 A)
  • TIC Relay Box Small Backing pump
  • Heater band & backing line isolation valve
  • Small backing pump
  • TIC Relay Box Inst & Small Backing pump
  • Changeover relays (3 off 250 V a.c. 3 A)
  • Heater band & backing line isolation valve
  • Small backing pump
  • Compatible mains backing pumps and accessories
    • Small backing pumps: E2M0.7 & 1.5, RV3, 5, 8 & 12, XDS5C & 10C, XDS5 & 10 & ESDP12
    • Heater band: BX70 & BX250
    • Backing line isolation valve: LCPV16EKA & LCPV25EKA

 


A&J Vacuum services and repairs Edwards Tic Instrument, along with all other Edwards Controller and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Edwards TIC Instrument Controller 200W, RS232/485, D39722000, With Support of Turbo Pump, Backing Pump and 3 Active Vacuum Gauges

Edwards TTIC D39722000 Instrument Technical Data:


Pump / TIC power 100W / 200W
  EXT75DX Fast / Fast
  nEXT 80W Slow / Slow
  nEXT 160W Slow / Fast
Mains input  
    Electrical supply 90 to 264 V a.c. 47 to 63 Hz
  Power consumption (max) 215 VA
  Peak inrush current 10.3 A @ 110 V a.c. / 23.0 A @ 230 V a.c.
  Earth stud M4
Auxiliary terminals  
  Air cooling fan 24 V d.c. 3 W max, ACX70, ACX75 & ACX250H
  Vent valve 24 V d.c. 2 W max, TAV5 & TAV6
Dimensions  
  Electronics housing 110 mm high x 105 mm wide x 245 mm deep
  Front panel 106 mm wide x 128 mm high
Weight 3.5 kg
Operating temp +0 ° to +40 °C
Storage temp -30 ° to +70 °C
Max ambient operating humidity 90% RH non-condensing at 40 °C
Max operating altitude 3000 m
Electronic Design EN 61010-1
Electromagnetic Compatibility EN 61326 Industrial Location, Class B Emissions
Enclosure rating IP20

 


A&J Vacuum services and repairs Edwards Tic Instrument, along with all other Edwards Controller and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.


Contact
address
A&J Vacuum Services
16 Somerset Place, Clifton, NJ 07012
customer service
Customer Service
Phone: 973-249-0854, Fax: 973-249-0855
working hours
Service Hours
8:00am to 5:00pm, Monday thru Friday EST