The Pfeiffer HLT560 SmartTest Helium Leak Detectors PTL02001 are microprocessor-controller leak detecting instruments. All the processes in the instrument are controlled automatically.
When searching for leaks with the SmartTest the test gas entering or escaping
through leaks in the sample is detected.
For gas to flow through a leak a pressure difference between the inside and outside
of the sample is necessary. For this either excess pressure or vacuum pressure is
generated inside the sample.
In the vacuum method test gas is blown against the wall of the evacuated sample
from the atmosphere side. It enters the sample at leaks and is fed to the leak
detector.
The sample must be vacuum pressure-proof.
The sensitivity stages
counterflow ⇒ Twin-Flow™ low ⇒ Twin-Flow™ high
are run through.
The detection limit is lower than in the sniffing method. The helium concentration
at the leak must be known in order to quantify the leak. The state of equilibrium
must be waited for.
In the sniffing method the test gas escaping from leaks in the sample into the
atmosphere is detected.
The sample must withstand the applied test pressure.
In operation with the sniffing probe a constant gas flow is sucked in from the
atmosphere. The helium proportion of the air (5.2 ppm) causes a leak rate display
of approx. 1×10-6 mbar l/s which can be eliminated by the ZERO function.
To detect a leak, the sniffing probe is applied to the points of the sample under
helium overpressure which are suspected of leaking. An increased leak rate value
indicates an increased concentration of helium and therefore a leak. The higher the
pressure and the helium concentration in the sample, the smaller the leaks which
can be detected.
The sensitivity stages
counterflow ⇒ Twin-Flow™ low
are run through.
The detection sensitivity and the quantifiability of the leak rate are less favourable
than in the vacuum pressure leak detection.
With its modular design and it's new wireless remote control RC 500 WL, the SmartTest offers the perfect solution for vacuum leak detection. Fast, reliable results, lowest detectable leak rates and short recovery time are what characterize the SmartTest and make it such a powerful and dependable product.
Utmost sensitivity in identifying even the smallest leaks. Quantification of the leaks. Outstanding reproducibility of the measurement. Helium test gas is non-toxic and non-flammable. Short measurement times
Click Here for Pfeiffer HLT 560 Manual
A&J Vacuum Services offers full repair/service of Pfeiffer HLT 560 SmartTest Helium Leak Detectors, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
Ambient temperature | 10-35 °C |
Cold start to ready | 3 min |
Detectable gases | 4He, 3He, H2 |
Flange (in) | DN 25 ISO-KF |
Inlet pressure max. | 25 mbar |
Input: Analog | ActiveLine vacuum transmitter |
Input: Digital | Start/Stop, Vent, Cero, Calibration, Bypass detection |
Interface | RS-232-C, RS-485 |
Leak rate display | 10-12 - 1 mbar l/s |
OuTPU t: Analog | Leakage rate signal 2 x 0-10 V, lin/log |
OuTPU t: Digital | Ready to start, Ready to measure, Leak, Error, Bypass valve, Calibration acknowledge |
Power consumption | 300 VA |
Pumping speed for He | 2.5 l/s |
Pumping time to ready: 0.5 l test volume | no |
Pumping time to ready: 0.5 l test volume | 1 s |
Pumping time to ready: 100 l test volume | no |
Pumping time to ready: 100 l test volume | 1300 s |
Pumping time to ready: 10 l test volume | no |
Pumping time to ready: 10 l test volume | 125 s |
Relay: Switching capacity | 2 x 230 V AC |
Relay: Switching current | 3 A |
Response time | 0.5 s |
Roughing pump | Yes |
Roughing pump | Single-stage rotary vane pump, oil sealed. Volume flow rate 4 m³/h at 50 Hz, 5.1 m³/h at 60 Hz |
Smallest detectable leak rate for He (according AVS 2.1) |
< 5·10-12 mbar l/s |
Sniffer probe, detectable leak rate | < 5·10-8 mbar l/s |
Supply | no |
Supply | 120 V, 60 Hz |
Test method | Vacuum and sniffing leak detection |
Weight | 44 kg |
Click Here for Pfeiffer HLT 560 Manual
A&J Vacuum Services offers full repair/service of Pfeiffer HLT 560 SmartTest Helium Leak Detectors, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.
The Pfeiffer HLT560 SmartTest Helium Leak Detectors PTL02001 are microprocessor-controller leak detecting instruments. All the processes in the instrument are controlled automatically.
When searching for leaks with the SmartTest the test gas entering or escaping
through leaks in the sample is detected.
For gas to flow through a leak a pressure difference between the inside and outside
of the sample is necessary. For this either excess pressure or vacuum pressure is
generated inside the sample.
In the vacuum method test gas is blown against the wall of the evacuated sample
from the atmosphere side. It enters the sample at leaks and is fed to the leak
detector.
The sample must be vacuum pressure-proof.
The sensitivity stages
counterflow ⇒ Twin-Flow™ low ⇒ Twin-Flow™ high
are run through.
The detection limit is lower than in the sniffing method. The helium concentration
at the leak must be known in order to quantify the leak. The state of equilibrium
must be waited for.
In the sniffing method the test gas escaping from leaks in the sample into the
atmosphere is detected.
The sample must withstand the applied test pressure.
In operation with the sniffing probe a constant gas flow is sucked in from the
atmosphere. The helium proportion of the air (5.2 ppm) causes a leak rate display
of approx. 1×10-6 mbar l/s which can be eliminated by the ZERO function.
To detect a leak, the sniffing probe is applied to the points of the sample under
helium overpressure which are suspected of leaking. An increased leak rate value
indicates an increased concentration of helium and therefore a leak. The higher the
pressure and the helium concentration in the sample, the smaller the leaks which
can be detected.
The sensitivity stages
counterflow ⇒ Twin-Flow™ low
are run through.
The detection sensitivity and the quantifiability of the leak rate are less favourable
than in the vacuum pressure leak detection.
With its modular design and it's new wireless remote control RC 500 WL, the SmartTest offers the perfect solution for vacuum leak detection. Fast, reliable results, lowest detectable leak rates and short recovery time are what characterize the SmartTest and make it such a powerful and dependable product.
Utmost sensitivity in identifying even the smallest leaks. Quantification of the leaks. Outstanding reproducibility of the measurement. Helium test gas is non-toxic and non-flammable. Short measurement times
Click Here for Pfeiffer HLT 560 Manual
A&J Vacuum Services offers full repair/service of Pfeiffer HLT 560 SmartTest Helium Leak Detectors, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
Ambient temperature | 10-35 °C |
Cold start to ready | 3 min |
Detectable gases | 4He, 3He, H2 |
Flange (in) | DN 25 ISO-KF |
Inlet pressure max. | 25 mbar |
Input: Analog | ActiveLine vacuum transmitter |
Input: Digital | Start/Stop, Vent, Cero, Calibration, Bypass detection |
Interface | RS-232-C, RS-485 |
Leak rate display | 10-12 - 1 mbar l/s |
OuTPU t: Analog | Leakage rate signal 2 x 0-10 V, lin/log |
OuTPU t: Digital | Ready to start, Ready to measure, Leak, Error, Bypass valve, Calibration acknowledge |
Power consumption | 300 VA |
Pumping speed for He | 2.5 l/s |
Pumping time to ready: 0.5 l test volume | no |
Pumping time to ready: 0.5 l test volume | 1 s |
Pumping time to ready: 100 l test volume | no |
Pumping time to ready: 100 l test volume | 1300 s |
Pumping time to ready: 10 l test volume | no |
Pumping time to ready: 10 l test volume | 125 s |
Relay: Switching capacity | 2 x 230 V AC |
Relay: Switching current | 3 A |
Response time | 0.5 s |
Roughing pump | Yes |
Roughing pump | Single-stage rotary vane pump, oil sealed. Volume flow rate 4 m³/h at 50 Hz, 5.1 m³/h at 60 Hz |
Smallest detectable leak rate for He (according AVS 2.1) |
< 5·10-12 mbar l/s |
Sniffer probe, detectable leak rate | < 5·10-8 mbar l/s |
Supply | no |
Supply | 120 V, 60 Hz |
Test method | Vacuum and sniffing leak detection |
Weight | 44 kg |
Click Here for Pfeiffer HLT 560 Manual
A&J Vacuum Services offers full repair/service of Pfeiffer HLT 560 SmartTest Helium Leak Detectors, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.