Edwards EPS 800 Power Supply, 48 V DC, 800 W, for Edwards nEXT 730D, nEXT 930D, nExt 1230D, B8J200819, EPS800
Item#: PN40781 Mfr Part#: B8J200819 Manufacturer: Edwards Vacuum
Condition: New In Stock
New Edwards EPS 800 Power Supply, 48 V DC, 800 W, for Edwards nEXT 730D, nEXT 930D, nExt 1230D, B8J200819, EPS800

Features:
  • Edwards Turbo Pump Power Supply
  • P/N B8J200819
  • 800 Watts
  • For nEXT730D / 930D and 1230 D

PRICE: $2,100.00
WEB SAVINGS: $205.00
YOU PAY: $1,895.00
Item#: PN40781 Mfr Part#: B8J200819
Edwards EPS 800 Power Supply, 48 V DC, 800 W, for Edwards nEXT 730D, nEXT 930D, nExt 1230D, B8J200819, EPS800
Item#: PN40781 Mfr Part#: B8J200819 Manufacturer: Edwards Vacuum

Edwards EPS 800 Power Supply, 48 V DC, 800 W, for Edwards nEXT 730D, nEXT 930D, nExt 1230D, B8J200819, EPS800

Edwards EPS800 Turbo Pump Power Supply B8J200819 Overview:


The Edwards Turbo Pump Power Supply EPS 800 is designed to power large line of Edwards turbo Pump. EPS 800 has an output of 48 V DC with max 17.1 Amps with 800 Watts.

Edwards EPS 800 fiits the following ppumps listed below

  • Edwards nEXT730D
  • Edwards nEXT830D
  • Edwards nEXT1230D

The Edwards EPS 800 Power Supplly can be mounted directly on the side of the pump with option mounting bracket P/N B8J200832. EPS 800 can also be used as bench top Turbo Pump Power Supplly.

The Edwards EPS 800 requires special Mains cable (sold seperately)

 

 


A&J Vacuum services and repairs Edwards Tic Instrument, along with all other Edwards Controller and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Edwards EPS 800 Power Supply, 48 V DC, 800 W, for Edwards nEXT 730D, nEXT 930D, nExt 1230D, B8J200819, EPS800

Edwards EPS800 Turbo Pump Power Supply B8J200819 Overview:


Mains Voltage 100 - 240 V AC +_ 10 % 50/60 Hz
Maximumm Power consumption 1100 VA
Efficiency 84 %
Leakage Current 2 mA / 230 V AC
AC Fuse 15 A Slow Blow
DC voltage Output 48 V +-5 %
Maximumm DC Current 10.65 A
DC Peak Current 17.1 A
Maximum Rated Power 800 Watts
Weight 4.9 Kg
Protection Type EN 60529 UL50E IP 54 Type 1
Maximum Operatin Altitude 2000 m above sea level
   
   
   
 
   
   
   
   
   
   
   

 


A&J Vacuum services and repairs Edwards Tic Instrument, along with all other Edwards Controller and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Name
Price
| Service and Repair
We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.

Overview

Edwards EPS 800 Power Supply, 48 V DC, 800 W, for Edwards nEXT 730D, nEXT 930D, nExt 1230D, B8J200819, EPS800
Item#: PN40781 Mfr Part#: B8J200819 Manufacturer: Edwards Vacuum

Edwards EPS 800 Power Supply, 48 V DC, 800 W, for Edwards nEXT 730D, nEXT 930D, nExt 1230D, B8J200819, EPS800

Edwards EPS800 Turbo Pump Power Supply B8J200819 Overview:


The Edwards Turbo Pump Power Supply EPS 800 is designed to power large line of Edwards turbo Pump. EPS 800 has an output of 48 V DC with max 17.1 Amps with 800 Watts.

Edwards EPS 800 fiits the following ppumps listed below

  • Edwards nEXT730D
  • Edwards nEXT830D
  • Edwards nEXT1230D

The Edwards EPS 800 Power Supplly can be mounted directly on the side of the pump with option mounting bracket P/N B8J200832. EPS 800 can also be used as bench top Turbo Pump Power Supplly.

The Edwards EPS 800 requires special Mains cable (sold seperately)

 

 


A&J Vacuum services and repairs Edwards Tic Instrument, along with all other Edwards Controller and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Edwards EPS 800 Power Supply, 48 V DC, 800 W, for Edwards nEXT 730D, nEXT 930D, nExt 1230D, B8J200819, EPS800

Edwards EPS800 Turbo Pump Power Supply B8J200819 Overview:


Mains Voltage 100 - 240 V AC +_ 10 % 50/60 Hz
Maximumm Power consumption 1100 VA
Efficiency 84 %
Leakage Current 2 mA / 230 V AC
AC Fuse 15 A Slow Blow
DC voltage Output 48 V +-5 %
Maximumm DC Current 10.65 A
DC Peak Current 17.1 A
Maximum Rated Power 800 Watts
Weight 4.9 Kg
Protection Type EN 60529 UL50E IP 54 Type 1
Maximum Operatin Altitude 2000 m above sea level
   
   
   
 
   
   
   
   
   
   
   

 


A&J Vacuum services and repairs Edwards Tic Instrument, along with all other Edwards Controller and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.


Contact
address
A&J Vacuum Services
16 Somerset Place, Clifton, NJ 07012
customer service
Customer Service
Phone: 973-249-0854, Fax: 973-249-0855
working hours
Service Hours
8:00am to 5:00pm, Monday thru Friday EST