- 50/60HZ
- Ultimate Pressure: <1.10-4 mbar
- Stainless Steel Sheet Casing
- Operating Fluid: Hydrocarbon Oil
The Pfeiffer Adixen 2063 H1 Pascal Vacuum Pump has a direct drive motor, making it very compact. An electrically insulated fold-away handle is used for easy carrying. The anit-suckback system ensures the tightness of the pump during accidental or voluntary shutdowns.
The Adixen 2063 pump comes with universal three-phase motor can be disassembled independently of the rest of the pump, without the need to drain the oil case. On the oil case, a vertical sight glass can be used to inspect the oil level easily when filling the tank and during the operation of the pump.
Oil casing is made of stainless steel sheet. All welds are checked with the helium leak detector.
A rotary seal near the motor works with two other seals. The additional seals form a sealed compartment immersed in oil. If one of the seals deteroriates, it will become apparent from the chance in oil level.
A&J Vacuum Rrepair / Rebuild and Rervice all Alcatel Adixen 2033H1 and 2063H1 Pascal Series Vacuum Pumps along with all other Alcatel Adixen Turbo Pump Service and repair. Please contact us for information on service program or click here http://www.ajvs.com/service.php to go to our Service page. We also offer Emergency service upon request. For more information Please contact A&J Vacuum Services at 973-249-0854.
| Frequency | 50/60 HZ |
| Number of stages | 2 |
| Nominal rotation speed | 1500/1800 tr/mn |
| Nominal flow rate | 65/78 m³/h 50 CFM |
| Flow rate Pneurop method | 12.5/15 m³/h 8.8 CFM |
| Partial ultimate pressure* with A120 oil |
<1.10-4 mbar <1.10-2 Pa |
| Exhaust pressure (maximum) | 2 bar |
| Internal gas volume | 3540 cm³ |
| Helium tightness ** - below atmospherique pressure - up to 1000 mbar |
≤1.10-7 mbar.l.s-1 <2.10-6 mbar.I.s-1 |
| Oil capacity (case) | 7.5 l |
| Oil capacity (sealed housing) | 0.27 l |
| Coling water flow rate | 2 l/mn |
| Water pressure | 3 bar |
| Weight (pump + motor) | 107 (236) kg (lb) |
A&J Vacuum Rrepair / Rebuild and Rervice all Alcatel Adixen 2033H1 and 2063H1 Pascal Series Vacuum Pumps along with all other Alcatel Adixen Turbo Pump Service and repair. Please contact us for information on service program or click here http://www.ajvs.com/service.php to go to our Service page. We also offer Emergency service upon request. For more information Please contact A&J Vacuum Services at 973-249-0854.
We can have a repair estimate ready for your equipment within the hour of submission.
Start your equipment repair process by creating a Service RMA.
A&J Vacuum Services provides repairs on all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair and Mass Spectrometer repair. Our highly trained technicians will deliver the highest quality service and satisfaction.
We operate at our own full service repair facility which allows us to quickly and economically perform all repairs in a timely manner. Our advandced machinery enables us to handle all facets of the service and repair process. AJVS utilizes industry standard vibration analysis equipment and advanced testing equipment to troubleshoot and minimize all equipment failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.