- PN A46211000
- Filter Material: Stainless Steel Mesh
- Inlet Port: NW 25
- Outlet Port: 1" BSP / NW 25
The MF100 and MF100AE filters have fine-pore filter elements and a stainless steel wire-mesh filter element inside the steel body. The MF100 and MF100AE have two fine-pore filter elements.
The MF100AE is designed for use with corrosive and toxic gases and vapours. It is leak tested to 1 x 10-4 mbar.ls-1 (1 x 10-2 Pa.ls-1) and the filter elements, sight-glasses and seals are resistant to chemical attack.
The exhaust gases from the pump enter the inlet of the filter and pass down through the filter elements which separate and trap the oil mist.
The exhaust gases then pass through the stainless steel wire mesh element which traps any large droplets of oil carried by the filtered gas. The oil-free exhaust gases are then discharged through the outlet. The trapped oil drains into the base of the body of the mist filter.
The mist filter has a sight-glass so that you can see the level of oil trapped in the mist filter. The sight-glass indicates the maximum level at which the filter operates efficiently.
The sight-glass in the MF100 mist filter is glued in position inside the body of the mist filter. The sightglass in the MF100AE mist filter is secured in position by a bezel and 'O' rings. If necessary, you can replace the sight-glass in the MF100AE.
The body has a drain-valve so that you can drain the trapped oil from the filter.
The spring-loaded pressure-relief valve prevents a dangerous pressure build-up inside the pump and the body if the filter becomes blocked.
Designed to be used with the following Edwards Rotary Vane Vacuum Pumps:
A&J Vacuum Services offers full repair/service of Edwards Dry Scroll Pumps, along with all other brands offered on our website. AJVS also services Edwards pump. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
| Filter Material: | Stainless Steel Mesh |
| Number of Filter Elements: | 2 |
| Inlet Port: | NW 25 |
| Outlet Port: | 1" BSP |
A&J Vacuum Services offers full repair/service of Edwards Dry Scroll Pumps, along with all other brands offered on our website. AJVS also services Edwards pump. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
We can have a repair estimate ready for your equipment within the hour of submission.
Start your equipment repair process by creating a Service RMA.
A&J Vacuum Services provides repairs on all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair and Mass Spectrometer repair. Our highly trained technicians will deliver the highest quality service and satisfaction.
We operate at our own full service repair facility which allows us to quickly and economically perform all repairs in a timely manner. Our advandced machinery enables us to handle all facets of the service and repair process. AJVS utilizes industry standard vibration analysis equipment and advanced testing equipment to troubleshoot and minimize all equipment failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.