Overview:
The electronic frequency converter TURBOTRONIK NT 151/361 is used to drive the turbomolecular pumps TURBOVAC150, 360, 150 CSV, 360 CSV, 150 V, 360 V,151, 361, 151 C, 361 C installed in SCIEX instruments.
The TURBOTRONIK NT 151/361 converts the single phase mains voltage into a three-phase AC voltage for driving the asynchronous motor of the connected turbomolecular pump.
The following operational states are indicated by LEDs:
A relay output is also available for the NORMAL signal.
After starting the pump will run up at maximum current. This is indicated by the ACCEL LED.
Having reached approx. 80% of the nominal speed the NORMAL LED comes on and the ACCEL LED goes off.
The motor current is continuously monitored and controlled. If the speed cannot be maintained even at the maximum current due to external influences, for example because of a greater gas supply, the speed will drop but the pump will continue to operate.
When the speed drops below 150 Hz approx. the LED FAIL on the converter lights up.
The outputs are short-circuit proof and protected against no load conditions.
The TURBOTRONIK recognizes which pump is connected.
The connection line between TURBOTRONIK and TURBOVAC may be up to 100 m long.
The TURBOTRONIK is contained within a housing, and can be used as a table top instrument or for installation in a 19� rack. Operation is by means of the START and STOP buttons on the front plate.
THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.
Leybold NT 151/361 SCIEX Turbotronik Controller, 230-240 V, 86400
Technical Data:
Mains Frequency: 50/60 Hz
Power Consumption During Acceleration: 680 VA
Continuous Max. Power Consumption: 480 VA
Power Consumption Briefly: 750 VA
Current Limiting: 5 A
Electrical Safety As To: IEC 1010 / EN 61010 / VDE 0411
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.
Overview:
The electronic frequency converter TURBOTRONIK NT 151/361 is used to drive the turbomolecular pumps TURBOVAC150, 360, 150 CSV, 360 CSV, 150 V, 360 V,151, 361, 151 C, 361 C installed in SCIEX instruments.
The TURBOTRONIK NT 151/361 converts the single phase mains voltage into a three-phase AC voltage for driving the asynchronous motor of the connected turbomolecular pump.
The following operational states are indicated by LEDs:
A relay output is also available for the NORMAL signal.
After starting the pump will run up at maximum current. This is indicated by the ACCEL LED.
Having reached approx. 80% of the nominal speed the NORMAL LED comes on and the ACCEL LED goes off.
The motor current is continuously monitored and controlled. If the speed cannot be maintained even at the maximum current due to external influences, for example because of a greater gas supply, the speed will drop but the pump will continue to operate.
When the speed drops below 150 Hz approx. the LED FAIL on the converter lights up.
The outputs are short-circuit proof and protected against no load conditions.
The TURBOTRONIK recognizes which pump is connected.
The connection line between TURBOTRONIK and TURBOVAC may be up to 100 m long.
The TURBOTRONIK is contained within a housing, and can be used as a table top instrument or for installation in a 19� rack. Operation is by means of the START and STOP buttons on the front plate.
THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED.
Leybold NT 151/361 SCIEX Turbotronik Controller, 230-240 V, 86400
Technical Data:
Mains Frequency: 50/60 Hz
Power Consumption During Acceleration: 680 VA
Continuous Max. Power Consumption: 480 VA
Power Consumption Briefly: 750 VA
Current Limiting: 5 A
Electrical Safety As To: IEC 1010 / EN 61010 / VDE 0411
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.