- Pressure range 1 x 10-3 mbar
- Compatible with all of Edwards Active Gauge Controllers
- Operating Temperature; +5 to +60 *C
- Corrosion Resistant
Edwards APG100 series Active Pirani vacuum gaugesare available in 2 models. The
APG100-XM is the standard model and measures to 10-3 mbar. The APG100-XLC is a
corrosion resistant version with measurement to 10-4 mbar.
Both Edwards APG100 gauges feature compact size for easy installation, a linear output, and a replaceable
sensor tube. The new gauges are compatible with all Edwards Vacuum TIC instrument
controllers and other active gauge controllers and displays. They are also CSA, C/US
approved as well as fully RoHS compliant due to their lead-free construction
Features and Benefits;
*THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED*

A&J Vacuum services and repairs Edwards Vacuum Measurement Transmitters, along with all other Edwards Vacuum and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.
| Mass | 85 g |
| Internal volume | 5 cm3 |
| Enclosure rating | IP40 |
| Measurement range (APG100-XM) | Atmosphere to 10-3 mbar |
| Measurement range (APG100-XLC) | Atmosphere to 10-4 mbar |
| Accuracy (APG100-XM) | Typically +/- 15% at <100 mbar |
| Accuracy (APG100-XLC) | Typically +/- 15% at <10 mbar |
| Maximum over-pressure | 10 bar absolute |
| Operating temp range | 5o to 60o C |
| Storage temp range | -30o to 70 oC |
| Bake-out with no electronics | 150 oC |
| Humidity | 80% RH up to 31 oC decreasing linearly |
| to 50% RH at 40 oC and above | |
| Maximum altitude | 3000 m |
| Filament temperature | 100 oC above ambient |
| Electrical supply voltage | 15 to 30 V d.c. nominal |
| 13.5 V d.c. minimum | |
| 32 V d.c maximum | |
| Power consumption | 1 W |
| Output signal | 0 to 10 V d.c. nominal |
| Set-point – open collector transistor | |
| Rating | 30 V d.c. 100 mA |
| Range of set-point | 1.8 to 9.2 V dc |
| Fixed hysteresis | 500 mV (1/2 decade) |
| Level setting resolution | 6 mV |
A&J Vacuum services and repairs Edwards Vacuum Measurement Transmitters, along with all other Edwards Vacuum and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.
We can have a repair estimate ready for your equipment within the hour of submission.
Start your equipment repair process by creating a Service RMA.
A&J Vacuum Services provides repairs on all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair and Mass Spectrometer repair. Our highly trained technicians will deliver the highest quality service and satisfaction.
We operate at our own full service repair facility which allows us to quickly and economically perform all repairs in a timely manner. Our advandced machinery enables us to handle all facets of the service and repair process. AJVS utilizes industry standard vibration analysis equipment and advanced testing equipment to troubleshoot and minimize all equipment failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.