- Nominal diameter DN 25 ISO-KF
- Measurement range 1 · 10-4 – 1.2 · 103 hPa
- Filament Tungsten
- Materials in contact with media Stainless steel, Tungsten, gold, silicon oxide, glass
The Pfeiffer Vacuum MPT200 Gauge has been designed for vacuum measurement of gases in a pressure range of 5 x 10-9 through 1000 mbar.
MPT200 Digital Pirani gaugemust not be used for measuring flammable or combustible gases in mixtures containing oxidants (e.g. atmospheric oxygen) within teh explosion range.
The MPT200 Digital Pirani gauge consists of two separate measurement systems (Pirani and cold cathode system) the signals of which are combined in such a way that one uniform measurement signal is output.

*THERE ARE NO RETURNS ON THIS PRODUCT IF PRODUCT IS OPENED OR USED*
A&J Vacuum services and repairs Pfeiffer Vacuum Measurement, along with all other Pfeiffer Vacuum and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.
| Accuracy: 1 · 10-8 – 2 · 10-3 hPa | ± 25 % |
| Accuracy: 100 – 1000 hPa | ± 50 % |
| Accuracy: 10 – 100 hPa | ± 30 % |
| Accuracy: 2 · 10-3 – 10 hPa | ± 10 % |
| Ambient temperature | 5-60 °C | 41-140 °F | 278-333 K |
| Anode | Molybdenum |
| Bakeout temperature at the flange | 180 °C | 356 °F | 453 K |
| Cold cathode sensor control | Cold Cathode sensor can be switched on and off via interface |
| Feature | Cold Cathode sensor can be switched on and off via interface |
| I/O interfaces | RS-485, DeviceNet |
| Materials in contact with media | Tungsten, stainless steel, nickel, molybdenum, glass |
| Measurement range | 5 · 10-9 – 1 · 103 hPa |
| Measuring cycle | 10 ms |
| Method of measurement | Pirani/Cold Cathode |
| Nominal diameter | DN 25 ISO-KF |
| Protection category | IP54 |
| Repeatability: 1 · 10-2 – 10 hPa | ± 2 % |
| Repeatability: 1 · 10-8 – 1 · 10-2hPa | ± 7 % |
| Seal | Metal |
| Sensor cable length min. | 100 m |
| Supply: power consumption | 4.5 W |
| Supply: Power consumption max. | 4.5 W |
| Temperature: Storage | -40-65 °C | -40-149 °F | 233-338 K |
| Weight | 555 g |
A&J Vacuum services and repairs Pfeiffer Vacuum Measurement, along with all other Pfeiffer Vacuum and Turbo Pumps available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.
We can have a repair estimate ready for your equipment within the hour of submission.
Start your equipment repair process by creating a Service RMA.
A&J Vacuum Services provides repairs on all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair and Mass Spectrometer repair. Our highly trained technicians will deliver the highest quality service and satisfaction.
We operate at our own full service repair facility which allows us to quickly and economically perform all repairs in a timely manner. Our advandced machinery enables us to handle all facets of the service and repair process. AJVS utilizes industry standard vibration analysis equipment and advanced testing equipment to troubleshoot and minimize all equipment failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.