phone icon
Sales and Service
855 882 2886
basket icon
0 Items
$0.00
Find us on:
- Agilent Technologies / Varian
- Stokes Vacuum
- AJVS
- Thread Sealants
- Mounting Kits
- Leak Detectors
- Shipping Crates
- Cooling Water Circulators
- Muffle Furnace
- Rotary Evaporators
- Cold Traps
Pfeiffer Adixen ASM 340 D, Helium Leak Detector Dry Version, Universal Voltage 90-240 V, 50/60 Hz, KSBA00A0MM9A, Standard 15 pins I/O, RS-232
Item#: PN41195 Mfr Part#: KSBA00A0MM9A Manufacturer: Pfeiffer Vacuum
Condition: Remanufactured In Stock
Remanufactured Pfeiffer Adixen ASM 340 D, Helium Leak Detector Dry Version, Universal Voltage 90-240 V, 50/60 Hz, KSBA00A0MM9A, Standard 15 pins I/O, RS-232

Features:
  • Detectable Gases: 4He, 3He, H2
  • Backing Pump Capacity: 3.4 m3/h
  • Pumping Speed for Helium: 2.5 l/s
  • 15 Pin I/O, RS-232

Price upon request
Item#: PN41195 Mfr Part#: KSBA00A0MM9A
Pfeiffer Adixen ASM 340 D, Helium Leak Detector Dry Version, Universal Voltage 90-240 V, 50/60 Hz, KSBA00A0MM9A, Standard 15 pins I/O, RS-232
Item#: PN41195 Mfr Part#: KSBA00A0MM9A Manufacturer: Pfeiffer Vacuum 12 months warranty

Pfeiffer Adixen ASM 340 D, Helium Leak Detector Dry Version, Universal Voltage 90-240 V, 50/60 Hz, KSBA00A0MM9A, Basic 15 pins I/O, RS-232

 

Demo Pfeiffer Adixen ASM 340 Dry-Version Helium Leak Detector - Product Overview:


The Pfeiffer ASM 340 Dry Leak Detector is a high performance and very easy operation. Pfeiffer ASM 340 D guarantees high performance in a vacuum and sniffing leak detection. Adixen ASM 340D has various applications in small production environments for maintenance. The ASM 340-D has an effective pumping system and is available in two different versions dry or conventional. ASM 340 Dry is ultra-fast response time, short recovery time, easy operation. Pfeiffer ASM 340 D is the ideal solution for testing every day.

The Pfeiffer Adixen ASM 340 D (Dry Version) is a fully automatic, multi-purpose, compact table top dry helium/hydrogen leak detector. It consists of a Pfeiffer hybrid turbo drag vacuum pump and integrated oil-free roughing/backing pump. The ASM 340 D is ideally suited for both repetitive and non-repetitive sequential testing applications requiring ultra-clean operations for both inboard vacuum testing and outboard sniffer testing.

The ASM 340 D has a high sensitivity of 5*E-12 atm. cc/sec helium for inboard vacuum testing and 5*E-9 atm. cc/sec helium for outboard sniffer testing. It has a high quantitative inlet port test pressure of up to 25 mbar and 100 mbar qualitative testing pressure for massive leak identification. A high helium pumping speed (2.5 l/s) allows for quick response and recovery time.

The ASM 340 D has a high resolution 3-mass 180-degree deflection mass spectrometer with a VHS preamplifier, which adds enhanced sensitivity, stability, and durability. The ASM 340 D has a high performance (53 l/s) Pfeiffer hybrid turbo molecular drag pump with an added superior compression ration and low maintenance operation. The leak detector features an integrated 2CFM oil-free roughing module for high performance testing and a universal power supply for global accommodations.

 

Pfeiffer Adixen ASM 340 Dry leak detector comes with the following parts:


  • Leak Detector
  • Tool Kit
  • Cover
  • Power Cable 115V
  • Power Cable 220V
  • ASM Software
  • Manuals
  • Brochure

 

Pfeiffer/Adixen ASM 340 Control panel:


Pfeiffer/Adixen ASM 340 Dry Front panel, KSBA00A2MM9A

 

A Brief Description of the Science and Processes Involved in Helium Leak Detection:


Leak detection utilizing helium has a complex history dating back to the first half of the twentieth century and is used in a variety of processes and industries. Helium is used as the preferred tracer gas for eliminating leaks for a variety of reasons, including its non-flammability, non-toxicity, and inexpensiveness relative to other gases. Likewise, its inert state and small atomic size (helium has an atomic mass unit of 4 - second only to hydrogen, which has an AMU of 2) allows the gas to pass through leaks with ease. Unlike hydrogen however, helium is stable and non-explosive, providing a better alternative to be used in leak detection processes.

Utilizing the beneficial qualities of helium, a Helium Leak Detector pumps molecules of the gas into the device's analyzer cell through the use of vacuum pumps. The analyzer cell works on the principle of mass spectroscopy and measures the gas's partial pressure, displayed on the device's spectrometer. A higher reading indicates that a leak may be present, as represented by the quantity and concentration of helium molecules that have passed through a given space, as analyzed by the device. The leak detector converts this partial pressure of molecules into a flow rate. Helium leak detectors are known to be extremely precise and exhibit unparalleled accuracy in detecting leaks.

 

Adixen/Pfeiffer ASM 340 Dry Features:


  • Oil-Free Pumping System, Backing Pump Capacity of 3.4 m^3/h
  • 2.5 l/s Helium Pumping Speed
  • Customizable and Detachable Operator Interface
  • Integrated SD Card for Data Processing
  • Fully Configurable I/O Interface (37 Pins D-Sub) to Fit to Customer Needs
  • Integrated Wireless Bluetooth Communication
  • Wired or Wireless Remote Control and Sniffer Probes Available as Accessories

 

Adixen/Pfeiffer ASM 340 Dry Inside:


Adixen/Pfeiffer ASM 340 Dry Inside, KSBA00A2MM9A

 

Adixen/Pfeiffer ASM 340 Dry Dimensions:


Adixen/Pfeiffer ASM 340 Dry Dimension, KSBA00A2MM9A

 


Click Here for Pfeiffer/Adixen ASM 340 ASM340 ASM-340 D Operating


A&J Vacuum services and repairs Pfeiffer Adixen ASM 340D (Dry Version) Helium Leak Detectors, along with all other helium leak detectors available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

This item is non-returnable/non-refundable

Pfeiffer Adixen ASM 340 D, Helium Leak Detector Dry Version, Universal Voltage 90-240 V, 50/60 Hz, KSBA00A0MM9A, Basic 15 pins I/O, RS-232

 

Demo, Like New ! Adixen/Pfeiffer ASM340 Dry Technical Data:


Backing Pump Capacity 3.4 m³/h | 2 cfm | 56.67 l/min
Detectable Gases 4He, 3He, H2
Dimensions (L x W x H) 547 x 375 x 393 mm | 21.54 x 14.76 x 15.47 inch
Flange Model (Inlet) DN 25 ISO-KF
I/O Interfaces Configurable 37 pins I/O
Minimum Detectable Leak Rate for Helium (Sniffing Leak Detection) 5 * 10^[-10] Pa m^3/s | 3.75 * 10^[-9] Torr l/s | 5 * 10^[-9] mbar l/s
Minimum Detectable Leak Rate for Helium (Vacuum Leak Detection) 5 * 10^[-13] Pa m³/s  
Noise Level 52 dB(A)
Operating Temperature (Hard Vacuum Test) 0 – 35 °C
Operating Temperature (Sniffing Test) 0 – 35 °C
Power Consumption Max. 600 W
Protection Category IP20
Pumping Speed for He 2.5 l/s
Sound Pressure Level 52 dB(A)
Start-Up Time (20°C) Without Calibration ~ 3 min
Supply 90 – 240 V, 50/60 Hz
Test Method Vacuum and Sniffing Leak Detection
Warranty: Year(s) 2
Weight 45 kg | 99.21 lb

Click Here for Pfeiffer/Adixen ASM 340 ASM340 ASM-340 D Operating


A&J Vacuum services and repairs Pfeiffer Adixen ASM 340D (Dry Version) Helium Leak Detectors, along with all other helium leak detectors available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Name
Price
$162.00
$273.00
>
Item#:
PN23631
Mfr Part#:
123264/123264
$599.00
$1,990.00
Name
Price
Item#:
PN06060
Mfr Part#:
BG449208-T/BG449208-T
$1,830.00
Item#:
PN06061
Mfr Part#:
BG449209-T/BG449209-T
$1,970.00
Item#:
PN17212
Mfr Part#:
112535/112535
$689.95
$2,150.00
| Service and Repair
We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.

Overview

Pfeiffer Adixen ASM 340 D, Helium Leak Detector Dry Version, Universal Voltage 90-240 V, 50/60 Hz, KSBA00A0MM9A, Standard 15 pins I/O, RS-232
Item#: PN41195 Mfr Part#: KSBA00A0MM9A Manufacturer: Pfeiffer Vacuum

Pfeiffer Adixen ASM 340 D, Helium Leak Detector Dry Version, Universal Voltage 90-240 V, 50/60 Hz, KSBA00A0MM9A, Basic 15 pins I/O, RS-232

 

Demo Pfeiffer Adixen ASM 340 Dry-Version Helium Leak Detector - Product Overview:


The Pfeiffer ASM 340 Dry Leak Detector is a high performance and very easy operation. Pfeiffer ASM 340 D guarantees high performance in a vacuum and sniffing leak detection. Adixen ASM 340D has various applications in small production environments for maintenance. The ASM 340-D has an effective pumping system and is available in two different versions dry or conventional. ASM 340 Dry is ultra-fast response time, short recovery time, easy operation. Pfeiffer ASM 340 D is the ideal solution for testing every day.

The Pfeiffer Adixen ASM 340 D (Dry Version) is a fully automatic, multi-purpose, compact table top dry helium/hydrogen leak detector. It consists of a Pfeiffer hybrid turbo drag vacuum pump and integrated oil-free roughing/backing pump. The ASM 340 D is ideally suited for both repetitive and non-repetitive sequential testing applications requiring ultra-clean operations for both inboard vacuum testing and outboard sniffer testing.

The ASM 340 D has a high sensitivity of 5*E-12 atm. cc/sec helium for inboard vacuum testing and 5*E-9 atm. cc/sec helium for outboard sniffer testing. It has a high quantitative inlet port test pressure of up to 25 mbar and 100 mbar qualitative testing pressure for massive leak identification. A high helium pumping speed (2.5 l/s) allows for quick response and recovery time.

The ASM 340 D has a high resolution 3-mass 180-degree deflection mass spectrometer with a VHS preamplifier, which adds enhanced sensitivity, stability, and durability. The ASM 340 D has a high performance (53 l/s) Pfeiffer hybrid turbo molecular drag pump with an added superior compression ration and low maintenance operation. The leak detector features an integrated 2CFM oil-free roughing module for high performance testing and a universal power supply for global accommodations.

 

Pfeiffer Adixen ASM 340 Dry leak detector comes with the following parts:


  • Leak Detector
  • Tool Kit
  • Cover
  • Power Cable 115V
  • Power Cable 220V
  • ASM Software
  • Manuals
  • Brochure

 

Pfeiffer/Adixen ASM 340 Control panel:


Pfeiffer/Adixen ASM 340 Dry Front panel, KSBA00A2MM9A

 

A Brief Description of the Science and Processes Involved in Helium Leak Detection:


Leak detection utilizing helium has a complex history dating back to the first half of the twentieth century and is used in a variety of processes and industries. Helium is used as the preferred tracer gas for eliminating leaks for a variety of reasons, including its non-flammability, non-toxicity, and inexpensiveness relative to other gases. Likewise, its inert state and small atomic size (helium has an atomic mass unit of 4 - second only to hydrogen, which has an AMU of 2) allows the gas to pass through leaks with ease. Unlike hydrogen however, helium is stable and non-explosive, providing a better alternative to be used in leak detection processes.

Utilizing the beneficial qualities of helium, a Helium Leak Detector pumps molecules of the gas into the device's analyzer cell through the use of vacuum pumps. The analyzer cell works on the principle of mass spectroscopy and measures the gas's partial pressure, displayed on the device's spectrometer. A higher reading indicates that a leak may be present, as represented by the quantity and concentration of helium molecules that have passed through a given space, as analyzed by the device. The leak detector converts this partial pressure of molecules into a flow rate. Helium leak detectors are known to be extremely precise and exhibit unparalleled accuracy in detecting leaks.

 

Adixen/Pfeiffer ASM 340 Dry Features:


  • Oil-Free Pumping System, Backing Pump Capacity of 3.4 m^3/h
  • 2.5 l/s Helium Pumping Speed
  • Customizable and Detachable Operator Interface
  • Integrated SD Card for Data Processing
  • Fully Configurable I/O Interface (37 Pins D-Sub) to Fit to Customer Needs
  • Integrated Wireless Bluetooth Communication
  • Wired or Wireless Remote Control and Sniffer Probes Available as Accessories

 

Adixen/Pfeiffer ASM 340 Dry Inside:


Adixen/Pfeiffer ASM 340 Dry Inside, KSBA00A2MM9A

 

Adixen/Pfeiffer ASM 340 Dry Dimensions:


Adixen/Pfeiffer ASM 340 Dry Dimension, KSBA00A2MM9A

 


Click Here for Pfeiffer/Adixen ASM 340 ASM340 ASM-340 D Operating


A&J Vacuum services and repairs Pfeiffer Adixen ASM 340D (Dry Version) Helium Leak Detectors, along with all other helium leak detectors available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Pfeiffer Adixen ASM 340 D, Helium Leak Detector Dry Version, Universal Voltage 90-240 V, 50/60 Hz, KSBA00A0MM9A, Basic 15 pins I/O, RS-232

 

Demo, Like New ! Adixen/Pfeiffer ASM340 Dry Technical Data:


Backing Pump Capacity 3.4 m³/h | 2 cfm | 56.67 l/min
Detectable Gases 4He, 3He, H2
Dimensions (L x W x H) 547 x 375 x 393 mm | 21.54 x 14.76 x 15.47 inch
Flange Model (Inlet) DN 25 ISO-KF
I/O Interfaces Configurable 37 pins I/O
Minimum Detectable Leak Rate for Helium (Sniffing Leak Detection) 5 * 10^[-10] Pa m^3/s | 3.75 * 10^[-9] Torr l/s | 5 * 10^[-9] mbar l/s
Minimum Detectable Leak Rate for Helium (Vacuum Leak Detection) 5 * 10^[-13] Pa m³/s  
Noise Level 52 dB(A)
Operating Temperature (Hard Vacuum Test) 0 – 35 °C
Operating Temperature (Sniffing Test) 0 – 35 °C
Power Consumption Max. 600 W
Protection Category IP20
Pumping Speed for He 2.5 l/s
Sound Pressure Level 52 dB(A)
Start-Up Time (20°C) Without Calibration ~ 3 min
Supply 90 – 240 V, 50/60 Hz
Test Method Vacuum and Sniffing Leak Detection
Warranty: Year(s) 2
Weight 45 kg | 99.21 lb

Click Here for Pfeiffer/Adixen ASM 340 ASM340 ASM-340 D Operating


A&J Vacuum services and repairs Pfeiffer Adixen ASM 340D (Dry Version) Helium Leak Detectors, along with all other helium leak detectors available on our website. Please contact us for additional information concerning our service program or click http://www.ajvs.com/service.php to access our Services page. We likewise offer emergency service and repair upon request. For more information, please contact A&J Vacuum Services at (973) 249-0854. Thank you for choosing AJVS as your supplier.

Name
Price
$273.00
Item#:
PN23631
Mfr Part#:
123264/123264
$599.00
$1,990.00
Name
Price
Item#:
PN06060
Mfr Part#:
BG449208-T/BG449208-T
$1,830.00
Item#:
PN06061
Mfr Part#:
BG449209-T/BG449209-T
$1,970.00
Item#:
PN17212
Mfr Part#:
112535/112535
$689.95
$2,150.00
We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.


Contact
address
A&J Vacuum Services
16 Somerset Place, Clifton, NJ 07012
customer service
Customer Service
Phone: 973-249-0854, Fax: 973-249-0855
working hours
Service Hours
8:00am to 5:00pm, Monday thru Friday EST
All trademarks, trade names or logons mentioned or used are the property of their respective owners.