The PhoeniXL is a leak detector for Helium or Hydrogen. This instrument may be used to detect the location and the size of leaks on objects under test in two different ways:
The PhoeniXL is to be used for leak detection only. It must not be used as a pumping system (esp. pumping aggressive or humid gases.)
Pumping condensable gases and steams: When pumping test sample water vapour that is inside the test object can attain to the fore pump. With the water vapor that is in the air - especially in humid areas or when using humid or wet test samples - the acceptable compatibility of water vapor or capacity of water vapor respectively can be exceeded. The steam in the oil of the pump condeses when the water vapor rises over the acceptable value. So the attribute of the oil changes and danger of corrosion occures for the pump.
While using the leak detector with condensable gases and steams the oil of the fore pump has to be controlled regularly. So you can recognize a condensation of water vapor in the pump. Usually the oil is light and lucent. When water vapor is inside it gets blear and milky at operating state temperature. When turning the pump off water vapor condensates and raises the part of water in the oil.
The leak detector must not directly be switched off after the process, in which condensable gases or steams are pumped, is finished. It must be running (at least 20 minutes) with opend gas ballast valve until the oil of the pump is freed from detached steam.
Click Here for Leybold PhoeniXL 300 Operating Instruction
Click Here for Leybold Oerlikon Helium Leak Detectors
Click Here for Leybold Leybold Product Brochure
A&J Vacuum Services offers full repair/service of Oerlikon / Leybold PhoeniXL 300, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
Max. Inlet Pressure: | 15 mbar |
Minimum Detectable Helium Leak Rates: |
Vacuum Mode: < 3 x 10-11 mbar l/s |
Sniffer Mode: < 1 x 10-7 mbar l/s |
|
Minimum Detectable Hydrogen Leak Rates: |
Vacuum Mode: < 1 x 10-8 mbar l/s |
Sniffer Mode: < 1 x 10-7 mbar l/s |
|
Measurement Range: | 11 decades |
Pumping Speed (Helium) At Inlet: |
Gross: .02 l/s |
Precision: .4 l/s |
|
Fine: > 2.5 l/s |
|
Inlet Port: | DN 25 KF |
Power Consumption: | 350 VA |
Click Here for Leybold PhoeniXL 300 Operating Instruction
Click Here for Leybold Oerlikon Helium Leak Detectors
Click Here for Leybold Leybold Product Brochure
A&J Vacuum Services offers full repair/service of Oerlikon / Leybold PhoeniXL 300, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.
The PhoeniXL is a leak detector for Helium or Hydrogen. This instrument may be used to detect the location and the size of leaks on objects under test in two different ways:
The PhoeniXL is to be used for leak detection only. It must not be used as a pumping system (esp. pumping aggressive or humid gases.)
Pumping condensable gases and steams: When pumping test sample water vapour that is inside the test object can attain to the fore pump. With the water vapor that is in the air - especially in humid areas or when using humid or wet test samples - the acceptable compatibility of water vapor or capacity of water vapor respectively can be exceeded. The steam in the oil of the pump condeses when the water vapor rises over the acceptable value. So the attribute of the oil changes and danger of corrosion occures for the pump.
While using the leak detector with condensable gases and steams the oil of the fore pump has to be controlled regularly. So you can recognize a condensation of water vapor in the pump. Usually the oil is light and lucent. When water vapor is inside it gets blear and milky at operating state temperature. When turning the pump off water vapor condensates and raises the part of water in the oil.
The leak detector must not directly be switched off after the process, in which condensable gases or steams are pumped, is finished. It must be running (at least 20 minutes) with opend gas ballast valve until the oil of the pump is freed from detached steam.
Click Here for Leybold PhoeniXL 300 Operating Instruction
Click Here for Leybold Oerlikon Helium Leak Detectors
Click Here for Leybold Leybold Product Brochure
A&J Vacuum Services offers full repair/service of Oerlikon / Leybold PhoeniXL 300, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
Max. Inlet Pressure: | 15 mbar |
Minimum Detectable Helium Leak Rates: |
Vacuum Mode: < 3 x 10-11 mbar l/s |
Sniffer Mode: < 1 x 10-7 mbar l/s |
|
Minimum Detectable Hydrogen Leak Rates: |
Vacuum Mode: < 1 x 10-8 mbar l/s |
Sniffer Mode: < 1 x 10-7 mbar l/s |
|
Measurement Range: | 11 decades |
Pumping Speed (Helium) At Inlet: |
Gross: .02 l/s |
Precision: .4 l/s |
|
Fine: > 2.5 l/s |
|
Inlet Port: | DN 25 KF |
Power Consumption: | 350 VA |
Click Here for Leybold PhoeniXL 300 Operating Instruction
Click Here for Leybold Oerlikon Helium Leak Detectors
Click Here for Leybold Leybold Product Brochure
A&J Vacuum Services offers full repair/service of Oerlikon / Leybold PhoeniXL 300, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.
A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.
We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.
We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.
Repair Steps
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.