Pfeiffer Adixen ASM 182 TD+ Standard Helium Leak Detector, 4.4 l/s Pumping Speed, 110/130 V, US Power Cable, D10RSS0PB710
Item#: PN17391 Mfr Part#: D10RSS0PB710 Manufacturer: Pfeiffer Vacuum
Condition: Discontinued
Discontinued Pfeiffer Vacuum Adixen ASM 182 TD+ Standard Helium Leak Detector, 4.4 l/s Pumping Speed, 110/130 V, US Power Cable, D10RSS0PB710

Features:
  • Frictionless ACP Dry Pump Featuring a Pumping Speed of 4.4 l/s
  • Backing Pump Capacity of 25 Cubic Meters/Hour
  • Includes English Language Voice Synthesizer
  • Remote Control Featuring Units in mbar I/s

This product has been discontinued and is no longer available for sale
Item#: PN17391 Mfr Part#: D10RSS0PB710
Pfeiffer Adixen ASM 182 TD+ Standard Helium Leak Detector, 4.4 l/s Pumping Speed, 110/130 V, US Power Cable, D10RSS0PB710
Item#: PN17391 Mfr Part#: D10RSS0PB710 Manufacturer: Pfeiffer Vacuum

Pfeiffer Vacuum Adixen ASM 182 TD+ Standard Helium Leak Detector, 4.4 l/s Pumping Speed, 110/130 V, US Power Cable, D10RSS0PB710

 

Pfeiffer Vacuum Adixen ASM 182 TD+ Helium Leak Detector - Product Overview:


The Pfeiffer Adixen ASM 182/192 series includes the latest evolutions of Adixen leak detectors, of which thousands are utilized in successful daily operation throughout the world in various applications and industries, including the automotive, chemical and mechanical, and electrical, electronics and packaging industries. A standout product of this series is the Adixen ASM 182 TD+ helium leak detector, utilized extensively in quality control purposes and maintenance of semiconductor processing systems, as well as quality control of vacuum components. Additionally, it is utilized as a blister-testing appliance within the pharmaceutical industry. This brand of Pfeiffer Adixen leak detectors offers several unique features, among them high sensitivity [5 x 10^(-12) mbar l/s], in addition to rapid response time while activated in spraying mode.

The Adixen ASM 182 TD+ leak detector is extremely flexible and user-friendly, with an emphasis on operational durability and ease-of-access. Automatic operation is provided throughout a vast range of processes, including atmospheric sniffer testing and refinement of extended gas lines and tubing. The ASM 182 TD+ mechanism likewise features a range of distinct operating modes, including gross leak mode and ultra-sensitive vacuum fine leak mode. Readily available basic functions, combined digital and analog display, and a range of advanced functions are also included for user convenience.

 

Pfeiffer/Adixen ASM 182 TD+ Features:


  • Frictionless ACP Dry Pump with 25 m^3/h Backing Capacity
  • 4.4 l/s Helium Pumping Speed
  • 110/130 V
  • US Power Cable (15 A)
  • English Language Voice Synthesizer
  • Remote Control Featuring Units in mbar l/s

 

Pfeiffer/Adixen ASM 182 TD+ Control Panel:


Pfeiffer/Adixen ASM 310 Front panel, BSAA0000MM9A

 

A Brief Description of the Science and Processes Involved in Helium Leak Detection:


Leak detection utilizing helium has a complex history dating back to the first half of the twentieth century and is used in a variety of processes and industries. Helium is used as the preferred tracer gas for eliminating leaks for a variety of reasons, including its non-flammability, non-toxicity, and inexpensiveness relative to other gases. Likewise, its inert state and small atomic size (helium has an atomic mass unit of 4 - second only to hydrogen, which has an AMU of 2) allows the gas to pass through leaks with ease. Unlike hydrogen however, helium is stable and non-explosive, providing a better alternative to be used in leak detection processes.

Utilizing the beneficial qualities of helium, a Helium Leak Detector pumps molecules of the gas into the device's analyzer cell through the use of vacuum pumps. The analyzer cell works on the principle of mass spectroscopy and measures the gas's partial pressure, displayed on the device's spectrometer. A higher reading indicates that a leak may be present, as represented by the quantity and concentration of helium molecules that have passed through a given space, as analyzed by the device. The leak detector converts this partial pressure of molecules into a flow rate. Helium leak detectors are known to be extremely precise and exhibit unparalleled accuracy in detecting leaks.

 

Pfeiffer/Adixen ASM 182 TD+ Dimensions (mm):


Pfeiffer/Adixen ASM 182 TD+, D10RSS0PB710


Click Here for Pfeiffer/Adixen ASM 182 TD+ Manual


A&J Vacuum offers full service and repair of Pfeiffer ASM 182 TD+ Helium Leak Detectors, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer Vacuum turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.

Pfeiffer Vacuum Adixen ASM 182 TD+ Standard Helium Leak Detector, 4.4 l/s Pumping Speed, 110/130 V, US Power Cable, D10RSS0PB710

 

Pfeiffer/Adixen ASM 182 TD+ Technical Data:


Backing Pump Oil-Free Backing Pump
Backing Pump Capacity 25 m³/h | 14.71 cfm | 416.67 l/min
Detectable Gases 4He
Dimensions (L x W x H) 601 x 487 x 456 mm | 23.66 x 19.17 x 17.95 inch
Flange Model (Inlet) DN 40 ISO-KF
I/O interfaces Digital inputs (Start, Vent, Calibration…); Digital outputs (Test modes, cycle in progress, helium signal above reject set point…); Analog outputs (Helium signal log, Inlet pressure)
Language English
Mains Cable Yes, US power cable  
Minimum Detectable Leak Rate for Helium (Sniffing Leak Detection) 1 * 10^[-8] Pa m3/s | 7.5 * 10^[-8] Torr l/s | 1 * 10^[-7] mbar l/s
Minimum Detectable Leak Rate for Helium (Vacuum Leak Detection) 5 * 10^[-13] Pa m³/s  
Noise Level 65 dB(A)
Power Consumption Max. 1500 W
Pumping Speed for He 4.4 l/s
Sound Pressure Level 65 dB(A)
Start-Up Time (20°C) With Calibration < 5.5 min
Start-Up Time (20°C) Without Calibration < 4 min
Supply 110 – 130 V, 50/60 Hz
Test Method Vacuum and Sniffing Leak Detection
Type High Performance - Compact
User Interface Basic User Interface
Weight 88 kg | 194 lb

 


Click Here for Pfeiffer/Adixen ASM 182 TD+ Manual


A&J Vacuum offers full service and repair of Pfeiffer ASM 182 TD+ Helium Leak Detectors, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer Vacuum turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.

| Service and Repair
We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.

Overview

Pfeiffer Adixen ASM 182 TD+ Standard Helium Leak Detector, 4.4 l/s Pumping Speed, 110/130 V, US Power Cable, D10RSS0PB710
Item#: PN17391 Mfr Part#: D10RSS0PB710 Manufacturer: Pfeiffer Vacuum

Pfeiffer Vacuum Adixen ASM 182 TD+ Standard Helium Leak Detector, 4.4 l/s Pumping Speed, 110/130 V, US Power Cable, D10RSS0PB710

 

Pfeiffer Vacuum Adixen ASM 182 TD+ Helium Leak Detector - Product Overview:


The Pfeiffer Adixen ASM 182/192 series includes the latest evolutions of Adixen leak detectors, of which thousands are utilized in successful daily operation throughout the world in various applications and industries, including the automotive, chemical and mechanical, and electrical, electronics and packaging industries. A standout product of this series is the Adixen ASM 182 TD+ helium leak detector, utilized extensively in quality control purposes and maintenance of semiconductor processing systems, as well as quality control of vacuum components. Additionally, it is utilized as a blister-testing appliance within the pharmaceutical industry. This brand of Pfeiffer Adixen leak detectors offers several unique features, among them high sensitivity [5 x 10^(-12) mbar l/s], in addition to rapid response time while activated in spraying mode.

The Adixen ASM 182 TD+ leak detector is extremely flexible and user-friendly, with an emphasis on operational durability and ease-of-access. Automatic operation is provided throughout a vast range of processes, including atmospheric sniffer testing and refinement of extended gas lines and tubing. The ASM 182 TD+ mechanism likewise features a range of distinct operating modes, including gross leak mode and ultra-sensitive vacuum fine leak mode. Readily available basic functions, combined digital and analog display, and a range of advanced functions are also included for user convenience.

 

Pfeiffer/Adixen ASM 182 TD+ Features:


  • Frictionless ACP Dry Pump with 25 m^3/h Backing Capacity
  • 4.4 l/s Helium Pumping Speed
  • 110/130 V
  • US Power Cable (15 A)
  • English Language Voice Synthesizer
  • Remote Control Featuring Units in mbar l/s

 

Pfeiffer/Adixen ASM 182 TD+ Control Panel:


Pfeiffer/Adixen ASM 310 Front panel, BSAA0000MM9A

 

A Brief Description of the Science and Processes Involved in Helium Leak Detection:


Leak detection utilizing helium has a complex history dating back to the first half of the twentieth century and is used in a variety of processes and industries. Helium is used as the preferred tracer gas for eliminating leaks for a variety of reasons, including its non-flammability, non-toxicity, and inexpensiveness relative to other gases. Likewise, its inert state and small atomic size (helium has an atomic mass unit of 4 - second only to hydrogen, which has an AMU of 2) allows the gas to pass through leaks with ease. Unlike hydrogen however, helium is stable and non-explosive, providing a better alternative to be used in leak detection processes.

Utilizing the beneficial qualities of helium, a Helium Leak Detector pumps molecules of the gas into the device's analyzer cell through the use of vacuum pumps. The analyzer cell works on the principle of mass spectroscopy and measures the gas's partial pressure, displayed on the device's spectrometer. A higher reading indicates that a leak may be present, as represented by the quantity and concentration of helium molecules that have passed through a given space, as analyzed by the device. The leak detector converts this partial pressure of molecules into a flow rate. Helium leak detectors are known to be extremely precise and exhibit unparalleled accuracy in detecting leaks.

 

Pfeiffer/Adixen ASM 182 TD+ Dimensions (mm):


Pfeiffer/Adixen ASM 182 TD+, D10RSS0PB710


Click Here for Pfeiffer/Adixen ASM 182 TD+ Manual


A&J Vacuum offers full service and repair of Pfeiffer ASM 182 TD+ Helium Leak Detectors, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer Vacuum turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.

Pfeiffer Vacuum Adixen ASM 182 TD+ Standard Helium Leak Detector, 4.4 l/s Pumping Speed, 110/130 V, US Power Cable, D10RSS0PB710

 

Pfeiffer/Adixen ASM 182 TD+ Technical Data:


Backing Pump Oil-Free Backing Pump
Backing Pump Capacity 25 m³/h | 14.71 cfm | 416.67 l/min
Detectable Gases 4He
Dimensions (L x W x H) 601 x 487 x 456 mm | 23.66 x 19.17 x 17.95 inch
Flange Model (Inlet) DN 40 ISO-KF
I/O interfaces Digital inputs (Start, Vent, Calibration…); Digital outputs (Test modes, cycle in progress, helium signal above reject set point…); Analog outputs (Helium signal log, Inlet pressure)
Language English
Mains Cable Yes, US power cable  
Minimum Detectable Leak Rate for Helium (Sniffing Leak Detection) 1 * 10^[-8] Pa m3/s | 7.5 * 10^[-8] Torr l/s | 1 * 10^[-7] mbar l/s
Minimum Detectable Leak Rate for Helium (Vacuum Leak Detection) 5 * 10^[-13] Pa m³/s  
Noise Level 65 dB(A)
Power Consumption Max. 1500 W
Pumping Speed for He 4.4 l/s
Sound Pressure Level 65 dB(A)
Start-Up Time (20°C) With Calibration < 5.5 min
Start-Up Time (20°C) Without Calibration < 4 min
Supply 110 – 130 V, 50/60 Hz
Test Method Vacuum and Sniffing Leak Detection
Type High Performance - Compact
User Interface Basic User Interface
Weight 88 kg | 194 lb

 


Click Here for Pfeiffer/Adixen ASM 182 TD+ Manual


A&J Vacuum offers full service and repair of Pfeiffer ASM 182 TD+ Helium Leak Detectors, along with all other brands of helium leak detectors offered on our website. AJVS also services Pfeiffer Vacuum turbo pumps and rotary vane pumps. Please contact us for information about our service and repair program or click http://www.ajvs.com/service.php to access our services page. We also offer emergency service upon request. For more information, please email A&J Vacuum Services at info@ajvs.com or call (973) 249-0854. Thank you for choosing AJVS as your supplier.

We can repair your:
  • Turbo Pumps
  • Rotary Vane Pumps
  • Dry Vacuum Pumps
  • Leak Detectors
  • Scroll Dry Pumps
  • Vacuum Boosters
  • Diaphragm Dry Pumps
You trusted worldwide leader of High Vacuum equipment!
Call us for your free repair estimate!
973-249-0854

A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service.

We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed.

We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies.

Repair Steps

  1. Full incoming inspection and run up time check, vibration check, drive inspection.
  2. Full disassembly, decontamination and cleaning.
  3. Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
  4. Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
  5. Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.

Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854.


Contact
address
A&J Vacuum Services
16 Somerset Place, Clifton, NJ 07012
customer service
Customer Service
Phone: 973-249-0854, Fax: 973-249-0855
working hours
Service Hours
8:00am to 5:00pm, Monday thru Friday EST